2020 Software Manual
98
Baud Rate
SW1
SW2
SW3
9600
On
Off
On
Table 5
DIP Switch Settings for Baud Rate
4. Switch 4 sets the data bits. Off = 7 Bits. On = 8 Bits.
Set the data bits to 8.
5. Use Switches 5 and 6 to set the parity to none.
Parity
SW5
SW6
None
Off
Off
Table 6
DIP Switch Settings for Parity
6. Switch 7 sets the Handshake mode. Off = Xon/Xoff.
On = DTR. Set the Handshake mode to Xon/Xoff.
7. Use switch 8 to set the data flow direction to serial to
parallel. Off = Serial to Parallel. On = Parallel to Serial.
8. Check your settings. It is imperative that you have the
correct settings or you will be unable to establish
communications. Your settings should match those in
Table 7
.
Содержание Photovac 2020
Страница 2: ......
Страница 3: ...Photovac Model 2020 Portable Photoionization Detector...
Страница 5: ...Reference No 600762 Rev E...
Страница 6: ......
Страница 13: ...Notices and Warnings...
Страница 14: ......
Страница 23: ...Introduction 1...
Страница 24: ......
Страница 38: ......
Страница 39: ...Tutorial Session 2...
Страница 40: ......
Страница 51: ...Tutorial Session 49 Mode on page 62 for details of switching between display modes...
Страница 60: ...2020 Software Manual 58 the datalogger or at the end of the datalogger will free usable space...
Страница 67: ...Tutorial Session 65...
Страница 68: ......
Страница 69: ...Detailed Operation 3...
Страница 70: ......
Страница 88: ...2020 Software Manual 86...
Страница 89: ...Connecting Accessories 4...
Страница 90: ......
Страница 115: ...Routine Maintenance 5...
Страница 116: ......
Страница 126: ...2020 Software Manual 124...
Страница 127: ...Troubleshooting 6...
Страница 128: ......
Страница 153: ...Troubleshooting 151...
Страница 154: ......
Страница 155: ...Technical Description 7...
Страница 156: ......
Страница 163: ...Appendices 8...
Страница 164: ......
Страница 181: ...Appendices 179...