N1015158 OPERATING MANUAL                              
 
 

2.3.5. External Control Possibilities

 

 

•  USB as standard. 

 
 

2.3.6. Dimensions 

 

Width

: 170 mm 

Height

: 280 mm 

Depth

: 430 mm 

Weight

: 6 kg 

 
 

2.3.7. Electrical data 

 

• Mains supply: 115/120 V or 220/240 V t 10% (factory setting 220/240 V) 
• Frequency: 50/60 Hz 
• Power: 120 VA 
• CE marked 

 
 

2.3.8. Description of Symbols 

 

Caution (Refer to accompanying documents). 
Symbol identifying the use of substances dangerous for health and environment. 
For these substances reference must be made to the laws and standards in force 
in the laboratory where this instrument is used. If using flammable or explosive 
substances   take the necessary precautions. 

 
Note: 

Connection  to  the  protective  ground  is  ensured  through  the  power  supply  cable 

which has to be in accordance with CEE 7 (IEC 23-5). 

 
 

2.3.9. Environmental conditions 
 

Performance and safety of the N1015158 are guaranteed at all times if it is operated under the 
following environmental conditions: 
 

• 

Installation category II with smaller transient over voltages (standard IEC 664) 

 
• 

Pollution degree 2 (standard IEC EN 61010-1) 

 
• 

The instrument is designed for internal use and altitudes below 2000 m 

 
• 

Mains voltage fluctuations must not exceed t10% of the nominal voltage 

 
• 

In locations where the relative humidity is lower than 80% 

 
• 

In locations free from dust or vapours of solvents and acids 

 

Содержание Peltier N1015158

Страница 1: ...rolled Fluid Circulator P N N1015158 Web address http las perkinelmer com OneSource Environmental directives htm Customer Care 1 800 762 4000 inside the USA 1 203 925 4602 outside the USA 0800 40 858...

Страница 2: ...1 Instrument description 4 2 2 N1015158 front and rear panel 5 2 3 Specifications 5 2 4 User interface 7 3 INSTALLATION 8 3 1 Checking the serial number 8 3 2 Checking main supply and fuses 8 3 3 Inst...

Страница 3: ...before starting any work inside the instrument Set the temperature to the required values on power on If using substances dangerous for health and the environment observe the laws and the standards i...

Страница 4: ...cording to the appropriate Perkin Elmer procedure e The equipment installation must be carried out in accordance with local requirements for responsibility and warranty 1 3 Safety makings CE mark The...

Страница 5: ...ferent metals The N1015158 provides temperatures ranging from 20 C to 60 C with 1 C stability and 0 1 C repeatability The features of the PCB 1500 include Peltier effect for heating and cooling the li...

Страница 6: ...et connections 2 3 Specifications 2 3 1 Keypad Display Unit The keypad display unit has a LCD screen All programmable parameters can be displayed in specific menus Three buttons are used to access the...

Страница 7: ...ed If using flammable or explosive substances take the necessary precautions Note Connection to the protective ground is ensured through the power supply cable which has to be in accordance with CEE 7...

Страница 8: ...The menu driven interface of the N1015158 allows the user to set the temperature and to view the configuration Three buttons are used to access the various functions and change the set parameters The...

Страница 9: ...inform Perkin Elmer immediately 3 2 Checking mains supply and fuses The N1015158 is factory set to operate at 220 VAC For operation at a voltage of 120 VAC the voltage selector at the rear of the cont...

Страница 10: ...sired value The new selected voltage will be now visible 3 3 Instrument set up 3 3 1 Start up The N1015158 is supplied completely assembled It is only necessary to make the appropriate connections Bef...

Страница 11: ...N1015158 has no special electrical external connections which would allow relay control etc Operations are performed either manually or via remote control using the USB connections Contact Perkin Elme...

Страница 12: ...it 1 only if the manual alarm reset option has been configured Menu page and functions of the buttons Menu Description L button Middle button R button Parameter User parameters Scroll Select Exit Conf...

Страница 13: ...equired to enter this menu CAUTION all these parameters are correctly set in the factory no changes should be made by the user Diagnostic parameters Description Unit of measure Cell temperature C Radi...

Страница 14: ...normal cleaning operations on the instrument only use a cloth dipped in water or neutral detergent do not use organic or abrasive solvents 5 2 Water level alarm Check for signs of water level It is re...

Страница 15: ...akage will be detected After the leakage is signalled on the N1015158 press the RESET ALARM button and remove the leakage immediately by checking all connections and fill up the system with distilled...

Страница 16: ...THE TANK WILL DROP AND WILL NEED TO BE TOPPED UP ONLY USE THE DISTILLED WATER SUPPLIED IN THE SEPARATE 200CC CONTAINER INSIDE THE PACKAGING TO TOP UP THE WATER IN THE UNIT Web address http las perkin...

Отзывы: