N1015158 OPERATING MANUAL                              

 

3.2.1. Changing the Mains Voltage Setting and Fuses 

 
The  voltage  set  is  shown  in  the  window  of  the  voltage  selector  below  the  mains  input 
connection. To change this setting switch off the unit at the mains switch, remove the mains 
cable and proceed as follows: 
 

• Slide out the fuse holder by inserting a  small screwdriver alternatively into the two 
slots at the left and right and pulling it out of the housing. 
 
• Remove  the  original  fuses  and  replace  with  those  of  correct  rating  (see  above)  and 
push the holder back in position. 

 

• Move the slider of the voltage selector with a screwdriver to the desired value. The 
new selected voltage will be now visible. 

 
 

3.3. Instrument set-up 
 
 
3.3.1. Start up 

 
The  N1015158  is  supplied  completely  assembled.  It  is  only  necessary  to  make  the 
appropriate connections. 
 
Before  starting  the  N1015158  you  should  familiarize  yourself  with  the  key  functions  as 
described in section 4. 
 

   
• Position the N1015158 at a suitable position in your system 
• Leave at least 5 cm in front and the back of the N1015158 for free air circulation. 
• Unscrew the tap on the "Vent" connector. 
• Mount the "Vent" connector supplied with the instrument. 
• Connect the supplied tubes to "Water in" and "Water out" fast connections. 
• Connect the tubes to the accessory to be thermostatted. 
• Switch on the N1015158. 
• The instrument initializes and after one minute it is ready to go. 
• Check the system for water level ("Water level alarm sensor). 
 

 
 
 
 
 
 
 
 
 

Содержание Peltier N1015158

Страница 1: ...rolled Fluid Circulator P N N1015158 Web address http las perkinelmer com OneSource Environmental directives htm Customer Care 1 800 762 4000 inside the USA 1 203 925 4602 outside the USA 0800 40 858...

Страница 2: ...1 Instrument description 4 2 2 N1015158 front and rear panel 5 2 3 Specifications 5 2 4 User interface 7 3 INSTALLATION 8 3 1 Checking the serial number 8 3 2 Checking main supply and fuses 8 3 3 Inst...

Страница 3: ...before starting any work inside the instrument Set the temperature to the required values on power on If using substances dangerous for health and the environment observe the laws and the standards i...

Страница 4: ...cording to the appropriate Perkin Elmer procedure e The equipment installation must be carried out in accordance with local requirements for responsibility and warranty 1 3 Safety makings CE mark The...

Страница 5: ...ferent metals The N1015158 provides temperatures ranging from 20 C to 60 C with 1 C stability and 0 1 C repeatability The features of the PCB 1500 include Peltier effect for heating and cooling the li...

Страница 6: ...et connections 2 3 Specifications 2 3 1 Keypad Display Unit The keypad display unit has a LCD screen All programmable parameters can be displayed in specific menus Three buttons are used to access the...

Страница 7: ...ed If using flammable or explosive substances take the necessary precautions Note Connection to the protective ground is ensured through the power supply cable which has to be in accordance with CEE 7...

Страница 8: ...The menu driven interface of the N1015158 allows the user to set the temperature and to view the configuration Three buttons are used to access the various functions and change the set parameters The...

Страница 9: ...inform Perkin Elmer immediately 3 2 Checking mains supply and fuses The N1015158 is factory set to operate at 220 VAC For operation at a voltage of 120 VAC the voltage selector at the rear of the cont...

Страница 10: ...sired value The new selected voltage will be now visible 3 3 Instrument set up 3 3 1 Start up The N1015158 is supplied completely assembled It is only necessary to make the appropriate connections Bef...

Страница 11: ...N1015158 has no special electrical external connections which would allow relay control etc Operations are performed either manually or via remote control using the USB connections Contact Perkin Elme...

Страница 12: ...it 1 only if the manual alarm reset option has been configured Menu page and functions of the buttons Menu Description L button Middle button R button Parameter User parameters Scroll Select Exit Conf...

Страница 13: ...equired to enter this menu CAUTION all these parameters are correctly set in the factory no changes should be made by the user Diagnostic parameters Description Unit of measure Cell temperature C Radi...

Страница 14: ...normal cleaning operations on the instrument only use a cloth dipped in water or neutral detergent do not use organic or abrasive solvents 5 2 Water level alarm Check for signs of water level It is re...

Страница 15: ...akage will be detected After the leakage is signalled on the N1015158 press the RESET ALARM button and remove the leakage immediately by checking all connections and fill up the system with distilled...

Страница 16: ...THE TANK WILL DROP AND WILL NEED TO BE TOPPED UP ONLY USE THE DISTILLED WATER SUPPLIED IN THE SEPARATE 200CC CONTAINER INSIDE THE PACKAGING TO TOP UP THE WATER IN THE UNIT Web address http las perkin...

Отзывы: