55
4.
When the subject is centered properly, click the Done button on the Manual Focus Window to
close the window. See Living Image
®
Software User Manual for IVIS Lumina XRMS Series III
(PN CLS137702RevB) for detailed instructions.
Glowing Materials
Always keep in mind that nearly EVERYTHING glows (that is, has the potential to phosphoresce
and contaminate the image). Most plastics, almost all tape, plants, paint, rodent food (mostly plants),
mouse urine, and animal bedding have been found to glow.
Use caution when introducing materials into the IVIS Lumina XRMS Series III. It is advisable to pre-
screen all items by imaging them alone, before imaging them with samples under study. PerkinElmer
recommends using non-powdered gloves when working with IVIS Lumina XRMS Series III
equipment.
7.7 System Shut Down Procedure
It is not recommended to power cycle the IVIS Lumina XRMS Series III (turning the system
components on and off). If it is necessary to shut down the imaging system for any reason, it is
important to follow the procedure below.
1.
Close the Living Image
software and save any information of interest at the prompt.
2.
Return the X-ray ON key selector switch to the OFF position and remove the key.
3.
Turn off the IVIS Lumina XRMS Series III imaging chamber using the power switch on the back
of the unit.
4.
Turn off the computer using the standard Windows
®
shut down procedure.
5.
Turn off the power to the other system components and power surge protection devices.
6.
If moving the system, unplug the devices from the wall.
If you have any problems during the shut down or start up procedure, please contact PerkinElmer
Technical Support for assistance (see
).
NOTE:
The Emergency OFF switch is not intended as a main X-ray source control
and should not be used to turn the X-ray function ON or OFF on a routine basis. It
should only be used in the unlikely situation where the X-ray source must be
immediately turned OFF. Under normal circumstances, it should be left in the ON