7. Operating the IVIS
®
Spectrum
36
Startup Procedure
The same procedure is used to restart the system whether the shut down was
intentional or due to a power outage.
1. Confirm that the rear panel switch is off, the front panel toggle switch is in
the down position (in standby), and the power cord is plugged into the
console and the wall socket.
2. Turn on the computer and monitor, and start the Living Image
®
software.
3. Turn on the main power switch on the rear panel.
— The status light is blue.
4. Turn on the front panel toggle switch by pulling the toggle switch out and
up.
— The status light is blue and blinking.
Figure 7.1
IVIS Spectrum
Table 7.1
IVIS Spectrum status light modes
Status Light Is...
Indicates...
Blue
The system is in standby mode.
Blue and blinking
The CCD camera cooling system is disabled due to a power
interruption. Use the Living Image software to initialize the
system and enable the cooling system.
Red
The system is initializing or acquiring an image.
Green
The system is ready.
Status light
Gas valve
Standby toggle switch
Bottom illumination
input optics
Содержание Caliper IVIS Spectrum
Страница 6: ...Contents iv ...
Страница 10: ...1 Welcome 4 This page intentionally blank ...
Страница 22: ...3 Warnings 16 This page intentionally blank ...
Страница 26: ...4 Legal Notices 20 This page intentionally left blank ...
Страница 30: ...5 Specifications 24 This page intentionally blank ...
Страница 40: ...6 IVIS Spectrum Components 34 This page intentionally blank ...
Страница 50: ...7 Operating the IVIS Spectrum 44 This page intentionally blank ...
Страница 68: ...10 Servicing Maintaining the ThermoCube Chiller Unit 62 This page intentionally blank ...
Страница 70: ...A Spare Parts List 64 This page intentionally blank ...
Страница 74: ...Index 68 ...