6.11 High Reflectance Hemisphere
You can special order a high reflectance hemisphere that is used to check for
light contamination in the imaging chamber
.
The hemisphere does not emit any photons when it is imaged in a light-tight and
contaminant-free imaging chamber. As a result, it is not visible in a luminescent
image. If a light leak or an internal phosphorescent contaminant is present in the
imaging chamber, the hemisphere reflects that light and is visible in a
luminescent image. If you suspect that there is light contamination in the
imaging chamber, see
Figure 6.10
High reflectance hemisphere