82 . STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide
Error Messages from the STA 6000
Error message ID
Error
Possible Cause Corrective
Action
ERR_DUE_TO_AUTOSAMPLER
AS 8000 entered
fatal error state.
This caused
STA 6000 to
enter fatal error
state
Refer to the AS
8000 error
message
Refer to AS 8000
troubleshooting
(above)
ERR_AUTOSAMPLER_RESPONS Timeout on
command.
Response from
command
transmitted to AS
8000
I2C connection
cable defective
Replace I2C cable
ERR_AUTOSAMPLER_I2C_WD
No link between
STA 6000 and AS
8000 via I2C
I2C connection
cable defective
Replace I2C cable
Содержание STA 6000
Страница 1: ...STA 6000 Installation and Hardware Guide THERMAL ANALYSIS ...
Страница 5: ...Introduction ...
Страница 7: ...Safety and Regulatory Information ...
Страница 17: ...Safety and Regulatory Information 17 ...
Страница 18: ......
Страница 19: ...Prepare the Laboratory ...
Страница 22: ...22 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...
Страница 23: ...Install Multiple Analyzers ...
Страница 26: ...26 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...
Страница 27: ...Install an STA 6000 ...
Страница 54: ...54 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...
Страница 55: ...STA 6000 Hardware ...
Страница 62: ...62 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...
Страница 63: ...AS 8000 Autosampler ...
Страница 84: ...84 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...