Install an STA 6000 . 39
To change the voltage:
WARNING
AVERTISSEMENT
Ensure that the mains pow er lead is disconnected from the
instrument before proceeding.
Assurez-vous que le cordon d'alimentation secteur est
débranché de l'instrument avant de continuer.
1.
Remove the fuse holder with a screwdriver by gently prying it out of the compartment.
2.
Once the holder is out far enough to hold onto, slide the fuse holder out of the slot.
3.
Remove the fuse (3.15 AT) from the 230 V position. This can be done using the point of
a pencil to gently lift the fuse out of the holder.
NOTE: It may be necessary to move the fuse slightly toward the back of the fuse pack
before prying it out of the holder.
Содержание STA 6000
Страница 1: ...STA 6000 Installation and Hardware Guide THERMAL ANALYSIS ...
Страница 5: ...Introduction ...
Страница 7: ...Safety and Regulatory Information ...
Страница 17: ...Safety and Regulatory Information 17 ...
Страница 18: ......
Страница 19: ...Prepare the Laboratory ...
Страница 22: ...22 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...
Страница 23: ...Install Multiple Analyzers ...
Страница 26: ...26 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...
Страница 27: ...Install an STA 6000 ...
Страница 54: ...54 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...
Страница 55: ...STA 6000 Hardware ...
Страница 62: ...62 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...
Страница 63: ...AS 8000 Autosampler ...
Страница 84: ...84 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...