132 . MultiScope System Microscope User's Reference
Electrical Connections
Fitting the Plug
The power cable for the electrical supply plugs into the back of the microscope. It
has a molded connector at one end. If it is necessary to fit a plug on the power
cable, use the wire color code below:
Plug Pin
Wire Color
(100/110/120 V)
Wire Color
(220/230/240 V)
Ground (Earth)
Green or Green/Yellow
Green/Yellow
Line
Black
Brown
Neutral
White
Blue
WARNING
To ensure safe and satisfactory operation of the microscope, it is
essential that the green or green/yellow ground (earth) wire of
the power cord is connected to a ground that complies with the
regulations of the local electricity supply authority (or equivalent
body); ground circuit continuity is essential for safe operation of
the equipment.
Содержание MultiScope System
Страница 1: ...MultiScope System Microscope User s Reference ...
Страница 5: ...Introduction ...
Страница 14: ...14 MultiScope System Microscope User s Reference ...
Страница 15: ...Warnings and Safety Information ...
Страница 26: ...26 MultiScope System Microscope User s Reference Warning Label Near the Lamp Housing Caution hot surface ...
Страница 31: ...Overview of the MultiScope System Microscope ...
Страница 44: ...44 MultiScope System Microscope User s Reference ...
Страница 45: ...Getting Ready to Use the Microscope ...
Страница 53: ...Tutorial Using the Microscope ...
Страница 62: ...62 MultiScope System Microscope User s Reference Figure 12 Reflectance Spectrum of the Coating on a Soda Can ...
Страница 63: ...Preparing Samples ...
Страница 83: ...Collecting Spectra with the Microscope ...
Страница 95: ...Operating the Optional Equipment ...
Страница 109: ...Applications ...
Страница 122: ...122 MultiScope System Microscope User s Reference ...
Страница 123: ...Maintenance ...