Instruction Manual peqPOWER 250 Volt Power Supply 

OPERATIONAL ELECTRICAL PARAMETERS 

Power Considerations 

Electrophoresis is the migration of a charged particle under the influence of an electrical field. The 
power supply output parameters voltage, current and power are related by the following two equations: 
 
Voltage (V) = Current (I) x Resistance (R); (V=I*R) 
Power (P) = Current (I) x Voltage (V); (P=I*V) 

Resistance 

Resistance of the assembled electrophoresis cell is dependent on the conductivity of the gel buffer, the 
thickness of the gel, and the number of gels being run. Although the resistance is determined by the gel 
system, the resistance can vary over the course of an electrophoretic separation. For instance, in the 
Tris-Glycine buffer system, the fast moving, highly conductive chloride ions in the gel are gradually re-
placed by the slower moving, less conductive glycine ions from the running buffer as the gel runs. As a 
result, the resistance of the gel increases as the chloride/glycine front moves down the gel, and the cur-
rent decreases. 

Voltage 

The velocity with which an ion moves in an electric field will vary in proportion to the field strength (volts 
per unit distance). The higher the voltage the faster an ion will move. 

Current 

Current is a function of the number of ions passing a given cross-section of the circuit at a given time. 
For a given gel/buffer system, at a given temperature, current will vary in proportion to the field 
strength (voltage) and/or cross-sectional area (number and/or thickness of the gels). Ions in solution 
and at a given voltage will move faster as the temperature increases, increasing current. 

Power 

The power in Watts, or the rate of heat generated by the system, is directly proportional to voltage and 
current (P=I*V). 
 

COMMON ERRORS FOUND WITH ELECTROPHORESIS POWER SUPPLIES 

No load 

 

The electrophoresis system is not connected to the power leads, check the power leads  

 

The electrophoresis system has a short, the Pt wire is broken or the banana connectors are dam-
aged 

 

Buffer concentration too low 

 

Buffer volume too low 

 

Short in power cord 

 

Current has dropped below acceptable rating (10 mA) 

 

PEQLAB_v0412E

 

12 

Содержание peqPOWER 250

Страница 1: ...Instruction Manual 250 Volt Power Supply 0412E Creating the future together...

Страница 2: ...TIONAL KEYS 6 GETTING STARTED 6 IMPORTANT GUIDELINES 7 OPERATIONAL MODES 8 CHOOSING LIMITING PARAMETER SETTINGS 10 TROUBLESHOOTING 11 OPERATIONAL ELECTRICAL PARAMETERS 12 COMMON ERRORS FOUND WITH ELEC...

Страница 3: ...with shielded banana plugs thus minimizing any potential shock hazard to the user Always use gel box sys tems that are compatible with the Power Supply have been designed for your specific application...

Страница 4: ...ase check carefully that the shipment is complete and has not been dam aged in transit For missing parts or to report any kind of damage please contact PEQLAB see TECH NICAL SERVICE AND ORDERING INFOR...

Страница 5: ...es Constant Voltage 1 V step Constant Current 1 mA step Constant Power 1 watt step Crossover Auto Display type LCD Graphic type Display size 53 64 x 15 64 mm W x H Pause function Yes Safety feature No...

Страница 6: ...tant Current and Constant Wattage Mode This manual describes the setup and operation of peqPOWER 250 Volt Power Supply including important information on safety and maintenance of the unit Features of...

Страница 7: ...Supply PEQLAB_v0412E 5 Description of Buttons and Switches Front View of peqPOWER 250 Volt Power Supply Rear View of peqPOWER 250 Volt Power Supply Control Buttons LCD Display Output Jacks Voltage Sel...

Страница 8: ...increase nu meric values STOP START PAUSE CONSTANT MODE GETTING STARTED Installing the peqPOWER 250 Volt Power Supply 1 Check the label located near the AC inlet to ensure that the unit is compatible...

Страница 9: ...sing the power switch on the rear of the in strument Upon start up the current settings appear on the screen Use the START PAUSE and STOP keys and output jacks for applications Use the CONSTANT key to...

Страница 10: ...e used during the entire duration of electrophoresis Review the guide lines provided in this manual before starting electrophoresis using the peqPOWER 250 Volt Power Sup ply A basic Constant Voltage C...

Страница 11: ...down using the key 2 When highlighted press Mode key 3 Select file number using the arrow keys 4 When file number is located press the Mode key 5 Select the number of steps by using the arrow keys 6...

Страница 12: ...ser extent continuous systems increase resistance during the run If you use the current limiting setting on the peqPOWER 250 Volt Power Supply the voltage will increase as resistance increases to sati...

Страница 13: ...e electrophoresis cell 5 Make sure buffers are prepared correctly and the recommended volume of buffer is added to the electrophoresis unit Operation stops with alarm The screen displays NO LOAD 6 Hig...

Страница 14: ...loride glycine front moves down the gel and the cur rent decreases Voltage The velocity with which an ion moves in an electric field will vary in proportion to the field strength volts per unit distan...

Страница 15: ...one parameter is limiting at a time The limiting parameter together with the conductivity in the electrophoresis system and the values for the other parameters determine the maximum output REPAIR AND...

Страница 16: ...Turn off the main power switch at the rear of the peqPOWER 250 Volt Power Supply and de tach the power cord from the rear of the peqPOWER 250 Volt Power Supply 2 Open the fuse compartment located insi...

Страница 17: ...er Cat No peqPOWER 300 V 230 V 4 300 500 mA 90 W 55 E300 230V peqPOWER 300 V 110 V 4 300 500 mA 90 W 55 E300 110V For technical questions and more detailed information on PEQLAB s products please visi...

Страница 18: ...RCHANTABILITY For research use only Not intended for any animal or human therapeutic or diagnostic use EQUIPMENT DISPOSAL This equipment is marked with the crossed out wheeled bin symbol to indicate t...

Страница 19: ...Instruction Manual peqPOWER 250 Volt Power Supply PEQLAB_v0412E 17 NOTES...

Страница 20: ...QLAB Biotechnologie GmbH 6404 Polling Tel 43 0 5238 84 169 info peqlab at www peqlab at UK PEQLAB Ltd Southampton SO31 7ZN Freecall UK 0808 20 21 302 info peqlab co uk www peqlab co uk USA PEQLAB LLC...

Отзывы: