ALARMS
GROUND LEAKAGE
A dangerous ground leakage is present (check thoroughly the
entire assembly). Your power supply is NOT defective! See Fre-
quently Asked Questions at the end of this manual.
*** ERROR ***
GROUND LEAKAGE
SHORT-CIRCUIT
Connected tank is short-circuited or has a very low resistance
(check thoroughly the complete assembly). Your power supply is
NOT defective! See Frequently Asked Questions at the end of
this manual.
*** ERROR ***
SHORT CIRCUIT
*** ERROR ***
OVERLOAD
OVERLOAD
Connected tank has a too low resistance (check specifications).
Your power supply is NOT defective! See Frequently Asked
Questions at the end of this manual.
*** ERROR ***
CURRENT < 1mA
CURRENT<1mA
The output current is lower than 1 mA (switch the alarm off in
the SET-UP menu). Your power supply is NOT defective! See
Frequently Asked Questions at the end of this manual.
*** ERROR ***
CALL SERVICE
CALL SERVICE
Failure in the electronics. Contact the PEQLAB technical service
(
r
s
You may also t y a general RESET: switch the instrument on while keeping the SET button pressed. Attention! All
stored program and data will be erased.
PEQLAB Biotechnologie GmbH_V1107
16