6. Service 

Note: 1) Refer to Photo-1, 2, Fig-1 for the location of the parts to be replaced. 

2) If any Numbers shown in the photo, they correspond to the numbers of the items listed below. 

6.1 Lamp (LP1, LP2) replacement 

A way of replacing the lamp bulb is described below. 

Caution: Just after turning off the lamp, it may be hot. Be careful not to get hest injury. 

1) Select the lamp that you want to replace with the lamp selector while the power of the unit is ON. 
2) Turn of the power of the unit and remove the lamp house lid by unscrewing the screw located on the 

upper center of the lid. 

3) Pull the lamp holder tab while lifting the ratchet until the lamp drops off. 
4) Remove the worn out lamp from the lamp socket and attach a new lamp to the socket. 

Caution: 1) Do not contaminate the lamp bulb with finger prints or something. It may cause a lamp trouble. 

2) Attach the lamp to the socket so that the protrusion on the lamp reflector edge faces inside. 

5) Put the lamp to the lamp holder so that the lamp reflector settles the hole in the holder, especially the 

protrusion meets the groove. 

6) Push the lamp holder tab until it latches. 
7) Close the lamp house with the lid by screwing the screw. 

Caution: A bar on the back of the lamp house lid is available, which turns off the interlock switch for the lamp 

ignition. Thus securely close the lid. 

 

 

6.2 Lamp harness (B510, 511) replacement 

The order of connecting the lamp harnesses to 
the terminal of the power supply unit must be kept 
as shown in Photo-4. 

1) Connect Lamp harness No.1 (B510/white) so 

the number attached to the wires “1” and “3” 
correspond to the terminal position “L1+” and 
“L1-“ as shown in Photo-4. 

2) Connect Lamp harness No.2 (B511/black) so 

the number attached to the wires “2” and “4” 
correspond to the terminal position “L2+” and 
“L2-“ as shown in Photo-4. 

 

3) Lift ratchet

3) Pull tab while 

lifting ratchet. 

5) Put lamp so reflector 

settles hole in holder 

5) Protrusion faces 

groove 

6) Push lamp 

holder tab 

Hole in lamp holder

Photo-3 Lamp replacement 

Lamp reflector

Lamp socket

For better referring inside, without cover 

Photo-4 Lamp harness

Referred by Photo-4 

Terminal

L1+

L1- 

L2- 

L2+ 

Wire number 
1, 2, 3, 4

Lamp harness 
No.1 

Lamp harness No.2

Содержание EPK-P

Страница 1: ...SERVICE MANUAL Version 00 High cost performance Video Processor PENTAX EPK p type C MNL 049 00 September 4 2009...

Страница 2: ...ew Parts list Notes for this Service Manual 1 This service manual must be handled as confidential Copy of the manual must be under your control In the case the maintenance contract expires the origina...

Страница 3: ...steps each RGB 9pin D sub x 1 Y C OUT S terminal x 2 Composite BNC connecter x 1 Output signal Analog output Y BNC connecter x 1 Remote Stereo mini plug x 2 Keyboard 6 pin Mini DIN x1 Control signal...

Страница 4: ...es The image can be mirrored or rotated and displayed on the monitor In the case that the item Image Direction on Display tab of Configuration Screen actuated by pressing F10 key is set with MIRROR th...

Страница 5: ...en the fan stops by accident it controls to shut down Lamp power supply PS4 6 DC power supply part Receives DC 12V and converts to 3 3V and 5V with the DC DC converter and the three terminal regulator...

Страница 6: ...1 LP1 Lamp 2 LP2 Lamp selector SW3 Motor MOT1 Lamp rotary disk Sensor board O700 Lamp ON OFF control Iris control 12VDC Error detection Calculated signal for iris Pump control Scope lock Lamp control...

Страница 7: ...p selector SW3 Lamp selector SW3 Iris unit MT1 Noise filter LF2 Front panel board F700 Fuse FU1 FU2 Mechanical block Integrated power supply unit PS4 Interlock switch SW2 Scope lock sensor PHC1 Proces...

Страница 8: ...arts The procedures of disassembling the mechanical and electrical parts are not described in detail in the service manual Unless otherwise mentioned detach from the outer parts to the inner parts or...

Страница 9: ...hat the lamp reflector settles the hole in the holder especially the protrusion meets the groove 6 Push the lamp holder tab until it latches 7 Close the lamp house with the lid by screwing the screw C...

Страница 10: ...front panel repeatedly until the brightness level becomes 5 6 Enter key command Ctrl J common between NTSC PAL to show the color bar chart 7 Adjust VR502 so that the edge of the iris plate just touche...

Страница 11: ...so that the air flow value indicated on the flow meter attains 2 5 0 1L min 7 Press PUMP POWER button repeatedly until the pump level High is selected 8 Press the up arrow key to select the pump leve...

Страница 12: ...d neutral terminals of the AC cable Point B Equipotential terminal of the EPK p Point C Equipotential wire of the dummy scope for dielectric strength test Point D Video signal BNC connector outer shel...

Страница 13: ...t A Point B AC input 4 Standard 1 Casing Leakage N C Connected Connected 110 VAC 100 A 1 2 N C 2 Connected Open 110 VAC 500 A 3 Protective earthing leakage S F C 3 Open Open 110 VAC 1000 A 4 N C Conne...

Страница 14: ...itch and Process board E700 Or the switch itself may be defective Lamp wire B510 511 Confirm if B510 White is connected to Lamp1 and B511 Black is connected to Lamp2 If they are connected the wrong wa...

Страница 15: ...ector Front panel board F700 Process board E700 Noise filter Power switch Integrated power supply unit PS1 Pump for Lamp Sensor board O700 Motor MOT1 Lamp selector SW3 CN9 CN3 CN1 Cable ID not availab...