1. Specifications 
 

Item Description 

Specification 

Voltage 120VAC 

230VAC 

Frequency 

50 - 60Hz 

Power consumption 

Less than 2.8A 

Less than 1.4A 

Power requirement 

Voltage fluctuation 

+/-10% 

Video system 

 

NTSC 

PAL 

Temperature 

10 – 40 

 

Relative humidity 

30 – 85 % (without condensation) 

Operation environment 

Atmospheric pressure   

700 – 1060 hPa 

Temperature 

-20 – 60 

 

Relative humidity 

0 – 95 % (without condensation) 

Storage environment 

Atmospheric pressure 

700 – 1060 hPa 

Lamp 

150W halogen lamp x2 

Lamp average life span 

50 hours 

Color temperature 

3,400K 

Iris control 

Automatic / Manual 

Illumination 

Brightness control 

Adjustable by +/- 5 steps each 

PENTAX Color Video 
Endoscope 

K-series (K-series with Y/C output) 

Scope compatibility 

PENTAX fiber scopes 

With use of appropriate fiber scope video adapter 

module. 

Color system 

 

Red and Blue each adjustable by +/- 5 steps each 

RGB 

9pin D-sub x 1 

Y/C OUT 

S terminal x 2 

Composite

BNC connecter x 1 

Output signal 

Analog 

output 

BNC connecter x 1 

Remote 

Stereo mini plug x 2 

Keyboard 6-pin 

Mini-DIN 

x1 

Control signal 

Serial signal 

USB x1 

Air pump system 

DC Diaphragm 

Pressure setting   

at flow rate at 0 

45 – 72 Kpa (maximum pressure) 

Air feeding system 

Standard air feed volume 
at inlet of the         

water bottle 

Low:  2.0 – 4.0 L/min 
High:  4.1 – 7.0 L/min 

Electric shock protection

Class – I equipment, 3 prong plug 

Degree of protection 
against electric shock 

BF type, using insulated endoscope 
Use on a heart is prohibited 

Classification electro 
medical equipment 

Degree of explosion 

proofing 

Do not use in potentially flammable surroundings. 

Electromagnetic 
Interference 

Electromagnetic 

compatibility 

EN60601-1-2(2001+A1:2006) for EU 
IEC60601-1-2 (2001+A1:2004) for other countries 

Medical electrical 
equipment general 
requirement for safety 

 

UL60601-1 IEC60601-1 

Size 

Dimensions 

380mm (W) x 430mm (D) x 155mm (H) 

Weight 

Main body 

11.0 Kg 

 

Table-1 Specifications 

Содержание EPK-P

Страница 1: ...SERVICE MANUAL Version 00 High cost performance Video Processor PENTAX EPK p type C MNL 049 00 September 4 2009...

Страница 2: ...ew Parts list Notes for this Service Manual 1 This service manual must be handled as confidential Copy of the manual must be under your control In the case the maintenance contract expires the origina...

Страница 3: ...steps each RGB 9pin D sub x 1 Y C OUT S terminal x 2 Composite BNC connecter x 1 Output signal Analog output Y BNC connecter x 1 Remote Stereo mini plug x 2 Keyboard 6 pin Mini DIN x1 Control signal...

Страница 4: ...es The image can be mirrored or rotated and displayed on the monitor In the case that the item Image Direction on Display tab of Configuration Screen actuated by pressing F10 key is set with MIRROR th...

Страница 5: ...en the fan stops by accident it controls to shut down Lamp power supply PS4 6 DC power supply part Receives DC 12V and converts to 3 3V and 5V with the DC DC converter and the three terminal regulator...

Страница 6: ...1 LP1 Lamp 2 LP2 Lamp selector SW3 Motor MOT1 Lamp rotary disk Sensor board O700 Lamp ON OFF control Iris control 12VDC Error detection Calculated signal for iris Pump control Scope lock Lamp control...

Страница 7: ...p selector SW3 Lamp selector SW3 Iris unit MT1 Noise filter LF2 Front panel board F700 Fuse FU1 FU2 Mechanical block Integrated power supply unit PS4 Interlock switch SW2 Scope lock sensor PHC1 Proces...

Страница 8: ...arts The procedures of disassembling the mechanical and electrical parts are not described in detail in the service manual Unless otherwise mentioned detach from the outer parts to the inner parts or...

Страница 9: ...hat the lamp reflector settles the hole in the holder especially the protrusion meets the groove 6 Push the lamp holder tab until it latches 7 Close the lamp house with the lid by screwing the screw C...

Страница 10: ...front panel repeatedly until the brightness level becomes 5 6 Enter key command Ctrl J common between NTSC PAL to show the color bar chart 7 Adjust VR502 so that the edge of the iris plate just touche...

Страница 11: ...so that the air flow value indicated on the flow meter attains 2 5 0 1L min 7 Press PUMP POWER button repeatedly until the pump level High is selected 8 Press the up arrow key to select the pump leve...

Страница 12: ...d neutral terminals of the AC cable Point B Equipotential terminal of the EPK p Point C Equipotential wire of the dummy scope for dielectric strength test Point D Video signal BNC connector outer shel...

Страница 13: ...t A Point B AC input 4 Standard 1 Casing Leakage N C Connected Connected 110 VAC 100 A 1 2 N C 2 Connected Open 110 VAC 500 A 3 Protective earthing leakage S F C 3 Open Open 110 VAC 1000 A 4 N C Conne...

Страница 14: ...itch and Process board E700 Or the switch itself may be defective Lamp wire B510 511 Confirm if B510 White is connected to Lamp1 and B511 Black is connected to Lamp2 If they are connected the wrong wa...

Страница 15: ...ector Front panel board F700 Process board E700 Noise filter Power switch Integrated power supply unit PS1 Pump for Lamp Sensor board O700 Motor MOT1 Lamp selector SW3 CN9 CN3 CN1 Cable ID not availab...