Pentax EPK-P Скачать руководство пользователя страница 8

 

- 6 -

 

 

5. Precaution - Important 

1) General 

You have to take utmost care to follow the general rules and practice for safety in servicing electrical 
medical equipment. 

2) Power off before servicing 

For electrical safety, make sure to disconnect the power cable before opening the cover. 

3) Route of electrical wire 

Be sure to keep the original routes of the electrical wires and the original positions of the cores.

 

Otherwise it may break the compatibility with the EMC regulation. You must put the wires back to the 
original routes after the maintenance. 

4) High temperature on the lamp 

Just after turning off the lamp it may be hot. Be careful not to suffer heat injuries when replacing a 
lamp or maintain parts around the lamp. 

5) Electrostatic Discharge 

To avoid the PCBs from getting damaged, wear anti-static protection such as a wristband and 
discharge the electrostatic charge before servicing. 

6) Disassembling parts 

The procedures of disassembling the mechanical and electrical parts are not described in detail in the 
service manual. Unless otherwise mentioned, detach from the outer parts to the inner parts or from the 
upper parts to the lower parts. In the case a special technique is necessary, it is described. You can 
refer to the exploded view pages as well. 

7) Optical unit mounting base 

It is made of PC+ABS (Polycarbonate, ABS resin) and parts as the lamp rotation disk are held with 
self-tapping screws. Thus tightening the self-tapping screw several times may screw up the mounting 
base, especially the screw holes in the base. Be careful not to spoil the screw holes by obliquely 
screwing the screws. If spoiling the screw hole it is necessary to remove the base with new one. 

8) Electrical safety test 

Be sure to do the following tests when you did a repair by opening the cove, after putting back all of 
the disassembled parts and the cover to the original position. The procedures are described in the 
following pages. 

a) Dielectric strength test, b) Protective earthing circuit resistance, c) Leakage current tests

 

Содержание EPK-P

Страница 1: ...SERVICE MANUAL Version 00 High cost performance Video Processor PENTAX EPK p type C MNL 049 00 September 4 2009...

Страница 2: ...ew Parts list Notes for this Service Manual 1 This service manual must be handled as confidential Copy of the manual must be under your control In the case the maintenance contract expires the origina...

Страница 3: ...steps each RGB 9pin D sub x 1 Y C OUT S terminal x 2 Composite BNC connecter x 1 Output signal Analog output Y BNC connecter x 1 Remote Stereo mini plug x 2 Keyboard 6 pin Mini DIN x1 Control signal...

Страница 4: ...es The image can be mirrored or rotated and displayed on the monitor In the case that the item Image Direction on Display tab of Configuration Screen actuated by pressing F10 key is set with MIRROR th...

Страница 5: ...en the fan stops by accident it controls to shut down Lamp power supply PS4 6 DC power supply part Receives DC 12V and converts to 3 3V and 5V with the DC DC converter and the three terminal regulator...

Страница 6: ...1 LP1 Lamp 2 LP2 Lamp selector SW3 Motor MOT1 Lamp rotary disk Sensor board O700 Lamp ON OFF control Iris control 12VDC Error detection Calculated signal for iris Pump control Scope lock Lamp control...

Страница 7: ...p selector SW3 Lamp selector SW3 Iris unit MT1 Noise filter LF2 Front panel board F700 Fuse FU1 FU2 Mechanical block Integrated power supply unit PS4 Interlock switch SW2 Scope lock sensor PHC1 Proces...

Страница 8: ...arts The procedures of disassembling the mechanical and electrical parts are not described in detail in the service manual Unless otherwise mentioned detach from the outer parts to the inner parts or...

Страница 9: ...hat the lamp reflector settles the hole in the holder especially the protrusion meets the groove 6 Push the lamp holder tab until it latches 7 Close the lamp house with the lid by screwing the screw C...

Страница 10: ...front panel repeatedly until the brightness level becomes 5 6 Enter key command Ctrl J common between NTSC PAL to show the color bar chart 7 Adjust VR502 so that the edge of the iris plate just touche...

Страница 11: ...so that the air flow value indicated on the flow meter attains 2 5 0 1L min 7 Press PUMP POWER button repeatedly until the pump level High is selected 8 Press the up arrow key to select the pump leve...

Страница 12: ...d neutral terminals of the AC cable Point B Equipotential terminal of the EPK p Point C Equipotential wire of the dummy scope for dielectric strength test Point D Video signal BNC connector outer shel...

Страница 13: ...t A Point B AC input 4 Standard 1 Casing Leakage N C Connected Connected 110 VAC 100 A 1 2 N C 2 Connected Open 110 VAC 500 A 3 Protective earthing leakage S F C 3 Open Open 110 VAC 1000 A 4 N C Conne...

Страница 14: ...itch and Process board E700 Or the switch itself may be defective Lamp wire B510 511 Confirm if B510 White is connected to Lamp1 and B511 Black is connected to Lamp2 If they are connected the wrong wa...

Страница 15: ...ector Front panel board F700 Process board E700 Noise filter Power switch Integrated power supply unit PS1 Pump for Lamp Sensor board O700 Motor MOT1 Lamp selector SW3 CN9 CN3 CN1 Cable ID not availab...