10. TROUBLE
SHOOTING
10.1 Trouble shooting ICM-2000
Trouble
Reason
Solution
Pressure drop
Upper needle leakage.
Replace O-ring (see paragraph
9.1).
Piercing leak.
Check rubber cone.
PET Bottle is pressed together
Use PET-bottle holder 005.218.
Values measured remain unstable Manometer leak (pos. 05).
Using a soap solution, check to
see if the screw thread
connection between manometer
and needle is leaking. If so,
clean the screw thread
connection and mount the
manometer on the needle with
LOCTITE 542.
Manometer (pos. 05) may be
defective.
Calibrate. If the manometer
shows too great a deviation or is
defective, it should be replaced.
10.2 Trouble shooting IAM
Trouble
Reason
Solution
IAM tap leaks
Tap has insufficient sealing.
Tighten the red nut on the tap.
Air volume measured is
improbably high
The lye is saturated (white)
and is no longer absorbing
CO
2.
Renew the lye.
Содержание HAFFMANS 006.625
Страница 2: ......
Страница 4: ......
Страница 6: ...HAFFMANS ICD Page 6 of 43 ICD...
Страница 8: ...HAFFMANS ICD Page 8 of 43 ICD...
Страница 10: ...HAFFMANS ICD Page 10 of 43 ICD...
Страница 12: ...HAFFMANS ICD Page 12 of 43 ICD...
Страница 16: ...HAFFMANS ICD Page 16 of 43 ICD...
Страница 22: ...HAFFMANS ICD Page 22 of 43 ICD...
Страница 24: ...HAFFMANS ICD Page 24 of 43 ICD...
Страница 26: ...HAFFMANS ICD Page 26 of 43 ICD...
Страница 28: ...HAFFMANS ICD Page 28 of 43 ICD...
Страница 30: ...HAFFMANS ICD Page 30 of 43 ICD...
Страница 32: ...HAFFMANS ICD Page 32 of 43 ICD...
Страница 34: ...HAFFMANS ICD Page 34 of 43 ICD...
Страница 36: ...HAFFMANS ICD Page 36 of 43 ICD...
Страница 40: ...HAFFMANS ICD Page 40 of 43 ICD...
Страница 42: ...HAFFMANS ICD Page 42 of 43 ICD...
Страница 44: ......
Страница 45: ......