Pentair HAFFMANS 006.625 Скачать руководство пользователя страница 20

 

HAFFMANS

  

ICD

 

 

 

Page 20 of 43 

ICD

 

There will now still be air in the hose between the IAM and the ICM-2000: 
 
This can be removed in three ways: 
 

 

Open the ICM-2000’s needle valve slowly until a drop of lye comes out of the needle. 
This will happen more quickly if the IAM is placed slightly higher than ICM-2000. Close 
the needle valve if all the air has been removed from the hose. Remove the drop of lye 
before starting the measurement. 

 

 

Pierce a bottle or can that does not need to be measured but can be used for testing. 
Shake to equilibrium and allow gas to escape from the can or bottle as small bubbles by 
opening the needle valve. Close the needle valve and wait until the lye has absorbed 
the CO

2

 in the hose, i.e., until the hose between the ICM-2000 and the IAM is 

completely filled with lye. Shaking the hose will speed this up. 

 

 

Remove the rubber cone from the ICM-2000. Connect a hose containing CO

2

 (0.5 bar 

maximum) to the ICM-2000’s needle. The hose may be warmed for ease of assembly. 
Open the needle valve slowly and expel the air from the hose using the CO

2

. Close the 

needle valve, remove the hose and assemble the pushbutton with the rubber cone. 

 
A non-return valve may be placed in the hose between the ICM-2000 and the IAM. 
The IAM is now filled with lye, without air, and is ready to be used for measurements. 
 

7.2 

Uses of the IAM 

The IAM has three possible uses, which are all based on the method given below.  
 

7.2.1  General IAM method 

 

1.  Slowly open the needle valve (pos. 8) and allow small bubbles of the head gas to bubble 

through the lye in the IAM. 

 

If the liquid does not foam up after opening the needle valve, the rear of the bottle must 
be tapped using a hammer (the foaming action forces the head gas out of the bottle). 

 
2.  Close the needle valve when gas no longer escapes from the bottle 

/can 

or

 as soon as liquid from the bottle/can enters the IAM’s hose, 

3.  Slide the leveller barrel (pos. 103) so that the liquid in the 

leveller 

barrel

 and the 

measuring burette

 (pos. 111) are at the same level 

(the hydrostatic pressure on the gas pocket will be neutralized and 
the correct volume will be indicated). 

4.  Open the sealing tap (pos. 112) and allow a small amount of lye to 

flow along the edge of the measuring chamber (this is now filled with 
air). 

5.  Adjust the liquid level in the 

leveller barrel

 and the 

measuring 

burette

 again so that they are the same. 

6.  Read the level on the IAM (the bottom of the meniscus) in ml. 
7.  Slide the leveller barrel upwards (maximum position) and open the 

sealing tap so that the air can escape from the measuring burette. 

8.  Close the sealing tap. 

Содержание HAFFMANS 006.625

Страница 1: ...WWW HAFFMANS NL HAFFMANS WWW HAFFMANS NL HAFFMANS INSTRUCTION MANUAL INPACK 2000 CO2 METER DIGITAL TYPE ICD...

Страница 2: ......

Страница 3: ...print or translation errors in this edition of the manual The specifications and design of this product are subject of change without notice All rights reserved No part of this publication may be rep...

Страница 4: ......

Страница 5: ...5 1 Unpacking 15 6 Operating the CO2 Meter 17 7 Operating the Air Meter 19 7 1 Preparations 19 7 2 Uses of the IAM 20 8 Calibration 23 9 Maintenance and cleaning 25 9 1 ICM 2000 maintenance semi annua...

Страница 6: ...HAFFMANS ICD Page 6 of 43 ICD...

Страница 7: ...diameter 120 mm Max bottom diameter 92 mm Dimensions Height 550 mm Width 170 mm Depth 250 mm Weight 2 4 kg 1 2 Inpack Air Meter Measuring rangeair content from 0 to 36 ml Subdivision from 0 to 6 ml 0...

Страница 8: ...HAFFMANS ICD Page 8 of 43 ICD...

Страница 9: ...lled bottles or cans for determination of the CO2 content The ICM 2000 can be used in combination with the Inpack Air Meter The Air Meter is intended exclusively for the measuring of the air content i...

Страница 10: ...HAFFMANS ICD Page 10 of 43 ICD...

Страница 11: ...ure to observe this information can result in major personal injury or loss of life 3 2 Safety instructions Check glass bottles beforehand for any cracks or breaks Use undamaged bottles only Before th...

Страница 12: ...HAFFMANS ICD Page 12 of 43 ICD...

Страница 13: ...etween gas and liquid in an enclosed space The law states that at a certain temperature the quantity of gas dissolved in a liquid is proportional to the partial pressure of the gas above the liquid 4...

Страница 14: ...head gas The air content of the head gas in a filled can or bottle can be determined by combining the ICC 2000 with the IAM The measuring method is mainly used to check and if necessary adjust the se...

Страница 15: ...ntact Haffmans BV or the representative agent designated by Haffmans B V immediately appendix III Always include the serial number order number or invoice number given by Haffmans BV of the ICM 2000 i...

Страница 16: ...HAFFMANS ICD Page 16 of 43 ICD...

Страница 17: ...hod 1 Bring the bottle or can according to the ASBC standard to a temperature of 25 0 0 2 C 77 0 0 4 F in a water bath This is not strictly necessary but recommended so that measurements can be reprod...

Страница 18: ...lay drops to about 0 5 bar and close the needle valve 6 Shaking bottle or can to equilibrium Shake the ICM 2000 with the pierced bottle or can and wait until the manometer value stops increasing 7 Rea...

Страница 19: ...e absorbed by 175 ml of lye solution To avoid pressure differences the liquid levels in both tubes must be made equal using the levelling barrel before taking the measurement The following must be car...

Страница 20: ...in the hose between the ICM 2000 and the IAM The IAM is now filled with lye without air and is ready to be used for measurements 7 2 Uses of the IAM The IAM has three possible uses which are all based...

Страница 21: ...the bottle or can Allow the head gas to escape and then shake the bottle or can immediately together with the ICM 2000 and then allow the gas to escape again Repeat this until no more gas is released...

Страница 22: ...HAFFMANS ICD Page 22 of 43 ICD...

Страница 23: ...sure 3 Compare the manometer pressure in terms of scale to the precision manometer pressure 4 If the ICM 2000 s manometer reading deviates by more than 1 from the reading on the precision manometer th...

Страница 24: ...HAFFMANS ICD Page 24 of 43 ICD...

Страница 25: ...le the sealing tap in the IAM fit the red disc first followed by the O ring and then the nut 9 3 ICM 2000 cleaning It is recommended to clean the ICM 2000 daily if measurements have been carried out M...

Страница 26: ...HAFFMANS ICD Page 26 of 43 ICD...

Страница 27: ...check to see if the screw thread connection between manometer and needle is leaking If so clean the screw thread connection and mount the manometer on the needle with LOCTITE 542 Manometer pos 05 may...

Страница 28: ...HAFFMANS ICD Page 28 of 43 ICD...

Страница 29: ...ates from the above standard the Haffmans CO2 formula should be adjusted for the constants A C and D Example using beer In beer the solubility of CO2 per 3 Plato changes by about 1 This means that the...

Страница 30: ...HAFFMANS ICD Page 30 of 43 ICD...

Страница 31: ...r their designated representative agent Always include the serial number order number or invoice number given by Haffmans BV of the ICM 2000 in any correspondence For answers to questions involving th...

Страница 32: ...HAFFMANS ICD Page 32 of 43 ICD...

Страница 33: ...ecified in the directions for use or which have been exposed to abnormal working conditions or abnormal handling The unlikely event that a defects occurs within the period covered by the guarantee the...

Страница 34: ...HAFFMANS ICD Page 34 of 43 ICD...

Страница 35: ...CONFORMITY We HAFFMANS B V Marinus Dammeweg 30 5928 PW Venlo Holland declare entirely under our own responsibility that the product CO2 METER TYPE ICM INPACK 2000 upon this statement has reference to...

Страница 36: ...HAFFMANS ICD Page 36 of 43 ICD...

Страница 37: ...6 05 Manometer 4 bar Manometer 6 bar 001 505 001 462 06 Manometer conductor port 006 376 07 Tube block housing 006 367 08 Needle valve 006 360 09 O ring 1 78 x 1 78 001 048 10 Seal 4 3 x 2 5 004 413 1...

Страница 38: ...37 Instrument base 005 025 38 Pan head screw M b 005 036 39 Sticker ICM 03 006 409 40 Hose socket 006 361 41 Clamping bush 4 x 17 006 668 42 Pan head screw with flange M 4x8 006 669 58 Charger R conne...

Страница 39: ...ase 5 gr O ring 3 68 x 1 78 36624 007 IAM INP 01 1324 Pos Description Art no 101 Brace 005 790 103 Levelling vessel 004 315 104 Connector nipple 004 727 105 Hose 6x4 mm 004 363 106 Hose 6x4 mm 004 363...

Страница 40: ...HAFFMANS ICD Page 40 of 43 ICD...

Страница 41: ...e rear position and close the needle valve 1 Preparing IAM Connect to the ICC Fill with lye and remove air from the hose 2 General method Open the needle valve and allow the head gas to bubble through...

Страница 42: ...HAFFMANS ICD Page 42 of 43 ICD...

Страница 43: ...HAFFMANS ICD ICD Page 43 of 43 APPENDIX VII DRAWINGS Description Drawing number Composition ICM 2000 INP 03 2061 Composition IAM INP 03 1102...

Страница 44: ......

Страница 45: ......

Отзывы: