Pull the electrical box “drawer” out far enough to be able to check the solenoid valves for
leaks when needed.
5.
Switch ON the tub’s circuit breaker, located on top of the electrical box. If water is flowing
from either the tub fill or the hand sprayer, turn it off by pressing the appropriate control
button once.
6.
Check each system (Tub fill, Hand shower, Disinfect Jets, and Rinse Jets) one at a time, for
leaks by turning it on and looking for leaks as it is flowing water. If no leak is found, turn off
the system. Note: The “Tub Fill” and “Hand Shower” are ON/OFF switches. The “Disinfect
Jets” and “Rinse Jets” buttons are momentary type switches and will only remain on as long
as you push and hold the buttons.
7.
If any leaks are found, tighten the connection to stop the leak.
8.
Open the tub drain and check the drain system for any leaks and fix if needed.
9.
Slide the electrical box “drawer” back into the tub and install the access door covers onto the
tub.
Finishing up:
There are a few adjustments that need to be made by the Penner Patient Care Representative.
Do
not operate this equipment until it is inspected by an authorized Penner Patient Care
Representative.
Warning!
Do not
attempt to operate this equipment before an authorized Penner Patient Care Representative
inspects it and selected personnel are trained in the operation of the equipment. The use of this
equipment before inspection, adjustment, and training by the Penner Representative could result
in damage to the equipment or injury to the operator and resident.