- 13 -
C
C
A
A
U
U
T
T
I
I
O
O
N
N
Some cleaners, disinfectants, and floor strippers contain
ingredients that are corrosive or abrasive. These solutions or
compounds may contain chlorine, acid, basic ingredients or
abrasives. DO NOT allow such solutions or compounds to
come in contact with your Penner equipment. Failure to heed
this caution could result in damage to the equipment and void
the warranty.
Cleaning of this equipment by untrained personnel could
result in injury to the operator or patient
1. Close and lock the door.
2. Press the Tub Fill Button and turn the Temperature Control
Knob all the way to the left to its warmest level to heat the
disinfectant solution and maximize its effectiveness.
3. Remove any visible tissue, residue, or fluids from the tub by
pressing the Shower Button and rinsing the inside tub surfaces
with the shower sprayer.
4. Press the Fill Button again to turn off the water. Allow the tub to
drain, and place the drain plug over the drain.
5. Press and hold the Disinfect Button located on the top of the tub
shown in (figure 1). As the button is held down, the properly
mixed cleaning solution is running through the air injection
system and out all of the air jets. Release the button after you
see solution coming out of all the air jets and you have 1 to 1
1/2 gallons of disinfectant solution in the foot well of the tub.