PressureMAT User Guide
Revision 8
Page 21/48
Pressure
Gives the option to display calculated values on the 4
th
line of the HOME screen. Options:
Normal:
Display pressure reading (PMAT4A or PMAT4R only)
Blank:
Display is blank (all models but PMAT2P, 2A, 3A, 2F & 3P)
TMP:
Displays (P1+P2)/2 - P3 (all models but PMAT1, 2P, 2A, 3A, 2F & 3P)
deltaP:
Displays (P1- P2) (all models but PMAT1, 2P, 2P, 2A, 3A, 2F, & 3P)
Alarm Latch
If turned on, alarm conditions will remain after an alarm state occurred but is no longer present
until the STOP button is pressed.
Backlight
Can be set to High, Medium or Low.
Audio Beep
Can be turned Off.
Keypad Secure
If turned On, the function to zero the sensor with the left arrow is disabled.
Communications
Displays setting not used in the unit at the present time. DO NOT CHANGE SETTINGS.
Date and Time
If SELECT/PROG button pressed, the following sub-menu appears:
DATE and TIME
> Date-Time DDMMMYY HH:MM:SS
Report Next DDMMMYY HH:MM:SS
Report Start DDMMMYY HH:MM:SS
Report Rate 000 hrs
Daylit Time Off
This menu is used mainly for controlling data output to a PC.
Note: See Appendix C on Data Collection to a PC
At this screen the UP/DOWN arrows are used to select the setting.