CONNECTIONS:
The external connections to the back panel are as shown below.
1.
POWER SUPPLY connected to the power inlet connector
2.
The PRESSURE SENSOR INPUTS connected via a DB15 connector
3.
The OUTPUT RELAY:
a.
NORMALLY OPEN- wired to terminals S and T via supplied connector (will switch to
CLOSED position with alarm condition)
b.
NORMALLY CLOSED- wired to terminals R and T via supplied connector (will
switch to OPEN position with alarm condition)
c.
POWER REMOVED- OPEN (S and T position) is the condition for relay(s)
4.
The 4-20mA ANALOG OUTPUTS are wired to terminals S and T via supplied connector as
shown
5.
Serial Port- send data RS232 to be collected by an external device (such as a PC)
6.
Auxiliary input- alternate power input connection for panel mounting
7.
Tare (S model only)- external relay (dry contact) closure across pins will tare sensor to zero