1.0 Specifications
Sensor Specifications
Sensor
Range
Response Time
H
2
S
0 – 100 ppm
T90 < 30 sec
CO
0 – 100 ppm
T90 < 30 sec
NH
3
0 – 50 ppm
T90 < 50 sec
O
2
0 – 25 %
T90 < 15 sec
SO
2
0 – 50 ppm
T90 < 120 sec
Contact Factory for other sensors.
Enclosure
Copper free Aluminum Instrument enclosure
CSA / UL / ATEX / IECEx Approved for Class I Division 1
Detector Housing
316 Stainless Steel
Weight & Dimension
4.5 Lbs ( 2Kg); 12”H x 6”W x 3.5”H (30.5cm x 15cm x 9cm)
Display
8 Character x 2 Line LCD Display
RF Frequency
900 MHz ISM Band, Frequency Hopping
OR
2.4
GHz ISM Band, frequency hopping wide band modulator
IEEE 802.15.4 Compliant
RF Certifications
FCC Part 15.247 USA
4214A 1208 Canada
ETSI Europe
(2.4 GHz Only)
Operating Range
1500 meters with dipole antenna line of sight
12 Km with High gain Yagi directional antenna
Power Input
7.2 Volts rechargeable Lithium Power Pak
Battery Charger
(purchased separately)
Universal Wall mount charger with 2.1 mm x 5.5 mm jack.
Rated Output: 12VDC @ 3 AH
Operating Temperature -40º F to +122º F (-40º C to +50º C) for H
2
S
(LT Series Detector Heads
)
-4º F to +140º F (-20º C to +60º C) for H
2
S
(EX Series Detector Heads)
-4º F to +131º F (-20º C to +55º C) for CO, SO2 & Oxygen
-40º F to +104º F (-40º C to +50º C) for Ammonia