T
ROUBLESHOOTING
Combination Warmer
▪
Operator’s Manual
▪
MN-47424
▪
Rev 1
▪
4/21
51
P131
P231
P331
P431
P531
P631
P731
Pad sensor 1
Pad sensor 2
Pad sensor 3
Pad sensor 4
Pad sensor 5
Pad sensor 6
Pad sensor 7
Heater pad over-temp error.
◾
Software disengages heating pads.
◾
Acknowledge error by pressing the standby button.
◾
Allow the warmer to cool.
◾
Contact service if the error persists.
*E-33
Cavity sensor
◾
Sensor reading is above maximum allowable temperature set-point and
over temp value. Blanket warmers trigger at 82°C (180°F). Fluid warmers
trigger at 71°C (160°F).
◾
Contact service if error persists.
*E-50
Analog to Digital
Converter Error
◾
Remove inventory, discard if necessary, and allow the warmer to cool
down.
◾
If error persists after cool down and reset, the control assembly should
be replaced by a qualified service technician. Contact service.
E-60
Real Time Clock
Checksum Error
(Blanket warmers only)
Real time clock rechargeable battery backup has discharged.
◾
Plug the warmer into the outlet for 30 minutes.
*E-61
Real Time Clock
(Blanket warmers only)
Real time clock not responding. Contact service if error persists.
E-62
Real Time Clock
(Blanket warmers only)
Timer overlay is present, but no real time clock is detected. Contact service.
*E-70
Pad Count Error
More heater pads detected than set for. Hold the standby button for 12
seconds until display shows "PAd#” (# = number of pads selected [3-7]).
Press up or down arrow to adjust to correct number of pads. [Blanket
warmers: 1.5ft
3
, 2.5ft
3
= 3 pads, 3.5ft
3
, 4.0ft
3
, 7.5ft
3
= 4 pads.
Fluid warmers: 2.5ft
3
& 4.0ft
3
= 3 pads].
*E-71
Personality Error
Contact service.
E-80
EEPROM Error
EEPROM not responding. Contact service if error persists.
*E-81
Calibration not
locked
Contact service.
*E-83
EEPROM Error
Contact service for help resetting the control.
E-87
EEPROM Error
Stored offsets corrupted. Offsets reset to 0.
◾
Control may need to be re-calibrated.
◾
Possible bad EEPROM.
◾
Contact service if error persists.
Code
Refers to
Action required
Содержание P-2420
Страница 4: ...FOREWORD Combination Warmer Operator s Manual MN 47424 Rev 1 4 21 4 This page intentionally left blank...
Страница 6: ...FOREWORD Combination Warmer Operator s Manual MN 47424 Rev 1 4 21 6 This page intentionally left blank...
Страница 10: ...SAFETY Combination Warmer Operator s Manual MN 47424 Rev 1 4 21 10 This page intentionally left blank...
Страница 24: ...SPECIFICATIONS Combination Warmer Operator s Manual MN 47424 Rev 1 4 21 24 This page intentionally left blank...
Страница 40: ...OPERATION Combination Warmer Operator s Manual MN 47424 Rev 1 4 21 40 This page intentionally left blank...
Страница 46: ...MAINTENANCE Combination Warmer Operator s Manual MN 47424 Rev 1 4 21 46 This page intentionally left blank...
Страница 54: ...TROUBLESHOOTING Combination Warmer Operator s Manual MN 47424 Rev 1 4 21 54 This page intentionally left blank...
Страница 60: ......
Страница 61: ......
Страница 62: ......
Страница 63: ......