M
AINTENANCE
Combination Blanket/Fluid Warmer
▪
Operator’s Manual
▪
M N-39930
▪
Rev 2
▪
4/20
40
Continued from previous page
Yearly
For yearly maintenance, do the following.
Check:
▪
the set-point temperature compared to the actual temperature displayed.
□
Check the cavity air temperature with a thermocouple placed 1” (25mm) from
the cavity sensor. Do not allow the thermocouple to touch any surface.
Monitor the temperature for approximately one hour in an empty cavity.
▪
the condition of the plug and cord and replace if damaged.
▪
the controller screen for excessive wear. Make sure the controller screen displays
and operates properly.
Clean:
▪
the shelves and interior of the warmer. See topic
How to Clean the Warmer
for the
list of approved cleaners.
Содержание P-2147
Страница 1: ...Operator s Manual Combination Blanket Fluid Warmer P 2147 P 2150 MN 39930 REV 02 4 20 EN...
Страница 58: ......
Страница 59: ......