Combination Blanket/Fluid Warmer
▪
Operator’s Manual
▪
MN-39930
▪
Rev 2
▪
4/20
19
P-2150
Built-In Cutout Specifications and Inlet Locations
Continued on next page
CAUTION:
Fire hazard.
Supply wires must be rated for a minimum wire temperature of
80°C.
NOTICE
Built-in units must be hardwired to the facility mains with a 20A breaker switch.
Breaker*
Wire Size**
Wire Temperature
20A
12 AWG (1mm
2
)
80°C (minimum)
*Electrical connections must meet applicable codes.
**IEC 60335 minimum.
Cutout Specifications
Inlet Options
Height
74.5” (1892.25mm)
1
Back of warmer
Width
29.25” (743mm)
2
Top of warmer
Depth
17” (432mm)
3
Left side of warmer
Top of unit clearance
12” (304.75mm)
—
—