background image

S

PECIFICATIONS

Combination Blanket/Fluid Warmer   

   Operator’s Manual   

   MN-39930   

   Rev 2   

   4/20

19

P-2150

 Built-In Cutout Specifications and Inlet Locations

Continued on next page

CAUTION: 

Fire hazard.

Supply wires must be rated for a minimum wire temperature of 

80°C.

  NOTICE  

Built-in units must be hardwired to the facility mains with a 20A breaker switch.

Breaker*

Wire Size**

Wire Temperature

20A

12 AWG (1mm

2

)

80°C (minimum)

*Electrical connections must meet applicable codes.

**IEC 60335 minimum.

Cutout Specifications

Inlet Options

Height

74.5” (1892.25mm)

1

Back of warmer

Width

29.25” (743mm)

2

Top of warmer

Depth

17” (432mm)

3

Left side of warmer

Top of unit clearance

12” (304.75mm)

Содержание P-2147

Страница 1: ...Operator s Manual Combination Blanket Fluid Warmer P 2147 P 2150 MN 39930 REV 02 4 20 EN...

Страница 2: ...not be reproduced or used in any manner whatsoever without the express written permission of Pedigo Products Inc Trademarks All trademarks referenced in this documentation are the property of their re...

Страница 3: ...pply the most durable convenient efficient and safe warming equipment on the market All warmers are manufactured and fully inspected in the USA with a commitment to quality Pedigo Repair Service Call...

Страница 4: ...FOREWORD Combination Blanket Fluid Warmer Operator s Manual MN 39930 Rev 2 4 20 4 This page intentionally left blank...

Страница 5: ...e Warmer for First Use 27 How to Turn On and Turn Off the Warmer 28 How to Operate the Blanket Warmer Bottom Cavity 30 How to Operate the TriMode Warmer Top Cavity 31 How to Change the Temperature Sca...

Страница 6: ...FOREWORD Combination Blanket Fluid Warmer Operator s Manual MN 39930 Rev 2 4 20 6 This page intentionally left blank...

Страница 7: ...below DANGER Danger indicates a hazardous situation which if not avoided will result in serious injury or death WARNING Warning indicates a hazardous situation which if not avoided could result in se...

Страница 8: ...e warmer Do not use this warmer for warming blood or blood products Do not cover or block any of the openings of this warmer Do not use this warmer in a wet location Only clean the warmer when the pow...

Страница 9: ...clean The warmer labels are legible Servicing the warmer Obey precautions in the manual on tags and on labels attached to or shipped with the warmer Only trained personnel are permitted to service or...

Страница 10: ...SAFETY Combination Blanket Fluid Warmer Operator s Manual MN 39930 Rev 2 4 20 10 This page intentionally left blank...

Страница 11: ...LABELS Combination Blanket Fluid Warmer Operator s Manual MN 39930 Rev 2 4 20 11 L ABELS Label Locations EC PHD 009217 P 2147 P 2150 Reach In P 2150 Pass Through Top View a c d b a b c d c d b...

Страница 12: ...150 Electrical warning label 3 Pedigo contact information 4 Quality inspected WARNING Hot surface label installed on the interior cavity wall S e c u r i t y S e a l S e c u r i t y S e a l S e c u r...

Страница 13: ...bility can only be achieved when the appliance is connected to an equivalent receptacle marked Hospital Grade EN61326 1 EN61000 3 2 EN61000 3 3 IP XO ordinary Model P 2147 P 2150 Intended use Bottom c...

Страница 14: ...P 2147 shown with feet EC DIM 004602 68 75 1743mm 68 10 1729mm 58 75 1492mm 29 25 743mm 56 1422mm 26 5 673mm 28 25 715mm 43 30 1100mm IEC Cord Inlet 23 75 603mm 23 75 603mm 30 763mm 15 381mm CAVITY I...

Страница 15: ...shown with casters EC DIM 004605 75 1901mm 68 1727mm 30 763mm 15 381mm CAVITY INSERT 26 5 673mm CAVITY 58 75 1492mm 29 25 743mm 56 1422mm 36 914mm CAVITY 35 890mm INSERT 26 5 673mm 28 5 721mm 49 5 12...

Страница 16: ...each In EC DIM 008782 15 381mm WITHOUT INSERT 36 914mm WITHOUT INSERT 34 75 882 75mm INSERT 59 5 1511mm 29 737mm 48 5 1232mm 19 75 501 75mm 21 25 539 75mm 72 75 1847 75mm IEC Cord Inlet 18 457mm CAVIT...

Страница 17: ...789 15 381mm WITHOUT INSERT 36 914mm WITHOUT INSERT 34 75 882 75mm INSERT 59 5 1511mm 29 737mm 19 75 501 75mm 21 25 539 75mm 72 75 1847 75mm IEC Cord Inlet 17 432mm CAVITY 17 432mm CAVITY 78 75 2000 2...

Страница 18: ...uilt In EC DIM 008794 15 381mm WITHOUT INSERT 36 914mm WITHOUT INSERT 34 75 882 75mm INSERT 59 5 1511mm 29 737mm 48 5 1232mm 19 75 501 75mm 21 25 539 75mm 72 75 1847 75mm IEC Cord Inlet 17 432mm CAVIT...

Страница 19: ...um wire temperature of 80 C NOTICE Built in units must be hardwired to the facility mains with a 20A breaker switch Breaker Wire Size Wire Temperature 20A 12 AWG 1mm2 80 C minimum Electrical connectio...

Страница 20: ...IFICATIONS Combination Blanket Fluid Warmer Operator s Manual MN 39930 Rev 2 4 20 20 Continued from previous page EC DIM 008799 1 2 3 74 5 1892 25mm 29 25 743mm 17 432mm 7 75 196mm 7 25 184 25mm 3 5 8...

Страница 21: ...on 1 Remove the box or crate Save all packing materials for inspection by the carrier 2 Cut and remove the retaining straps and plastic wrap 3 Remove the warmer from the pallet using an appropriate li...

Страница 22: ...ed from previous page P 2150 only 5 Open the accessories box and install the shelf clips and shelves into the cavities Result The warmer is now unpacked NOTE Set aside the toe panel s and screws The t...

Страница 23: ...the warmer Requirements The warmer must be installed on a level surface The warmer must not be installed in any area where it may be affected by dripping water high temperature or any other severely...

Страница 24: ...idelines shown below Refer to topic Specification Information for clearance requirements You reference the built in specification page for cutout dimensions P 2150 only 2 Move the warmer to the instal...

Страница 25: ...ge 5 Install the toe panel Connect power To connect electric power to the warmer do the following 6 Connect the plug to the electrical outlet Result The warmer is now installed and ready to be used EC...

Страница 26: ...INSTALLATION Combination Blanket Fluid Warmer Operator s Manual MN 39930 Rev 2 4 20 26 This page intentionally left blank...

Страница 27: ...eaner if stainless steel panels with an approved cleaning agent if painted panels 3 Clean the interior of the warmer with a damp cloth or approved cleaning agent 4 Dry the interior and exterior of the...

Страница 28: ...do the following Step Action 1 Set the power switch a to the ON l position 2 Press the standby button The screen turns on The blanket warmer and TriMode warmer screens must be turned ON separately The...

Страница 29: ...0 Rev 2 4 20 29 Continued from previous page Turning off the warmer To turn off the warmer do the following 3 Press and hold the standby button until the screen turns off The blanket warmer and TriMod...

Страница 30: ...mperature will display 4 Rotate the blankets daily from the bottom to the top of the load to ensure equal heating and to prevent discoloration Result The blankets are now warming WARNING Personal inju...

Страница 31: ...n will not illuminate Continued on next page WARNING Personal injury hazard Do not operate the warmer in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide in oxygen enric...

Страница 32: ...warming process 4 Press the temperature recall button to view the measured cavity temperature The measured cavity temperature will display for five seconds Then the set point temperature will display...

Страница 33: ...buttons to toggle between the temperature scales 4 Press the standby button to turn the screen on Result The temperature scale has now been changed How to Change the Sound Settings Before you begin Th...

Страница 34: ...do the following Step Action 1 Press and while holding the standby button press the up arrow button The lock indicator will illuminate Unlocking the controller To unlock the controller do the followi...

Страница 35: ...s of data The download time for one week of data should be under five seconds The warmer records various performance parameters during the following conditions automatically every five minutes while t...

Страница 36: ...e from the port 4 Close the blanket warmer door Result The data has now been downloaded Viewing the temperature data To view the temperature data do the following 5 Plug the USB drive into a computer...

Страница 37: ...Sho rtErrors 0x0c 12 19 2018 14 48 top off closed off 71 2 104 Diagnostics SensorSho rtErrors 0x0c 12 19 2018 14 53 bottom on closed on 71 7 130 Diagnostics SensorSho rtErrors 0x0c 12 19 2018 14 53 to...

Страница 38: ...OPERATION Combination Blanket Fluid Warmer Operator s Manual MN 39930 Rev 2 4 20 38 This page intentionally left blank...

Страница 39: ...maintenance do the following Check the door gasket for tears and holes Make sure that it is firmly attached to the door Check the seal when the door is closed the guards around the air temperature se...

Страница 40: ...mperature with a thermocouple placed 1 25mm from the cavity sensor Do not allow the thermocouple to touch any surface Monitor the temperature for approximately one hour in an empty cavity the conditio...

Страница 41: ...the warmer with a damp cloth or approved cleaning agent 5 Dry the interior and exterior of the warmer with a clean lint free cloth Leave the door open until the interior of the warmer completely drie...

Страница 42: ...cavity 4 Clean the shelves and interior with a damp cloth or approved cleaning agent 5 Dry the interior of the warmer with a clean lint free cloth Leave the door open until the interior of the warmer...

Страница 43: ...id Warmer Operator s Manual MN 39930 Rev 2 4 20 43 Warmer Disposal Decommissioning This product and its accessories must be disposed of according to local laws and regulations Do not dispose of this p...

Страница 44: ...MAINTENANCE Combination Blanket Fluid Warmer Operator s Manual MN 39930 Rev 2 4 20 44 This page intentionally left blank...

Страница 45: ...flashes The controller stores all settings and continues operation using these settings in the event of a power interruption Procedure To continue operation of the warmer do the following Step Action...

Страница 46: ...roller sounds an alarm when the warmer malfunctions Procedure If the alarm indicator light flashes do the following Step Action 1 Press the standby button to acknowledge and mute the alarm The alarm i...

Страница 47: ...f the warmer to the off O position and allowing the warmer to cool NOTICE Do not attempt to repair or service the warmer beyond this point Contact the manufacturer for the nearest authorized service a...

Страница 48: ...nsor from the warmer control PCB and use an Ohm meter to measure the resistance of the sensor Sensor PR 37140 should measure 10 KOhm 1 5 KOhm at 25 C 77 F Sensor SN 33541 should measure 100 Ohm 10 Ohm...

Страница 49: ...ocked Contact service E 83 EEPROM Error Contact service for help resetting the control E 87 EEPROM Error Stored offsets corrupted Offsets reset to 0 Control may need to be re calibrated Possible bad E...

Страница 50: ...Fan or Fan Sensor Failure Check to make sure the fan sensor wires did not disconnect from the fan or the control board If the error persists after checking the wires replace the fan Red iv fluid temp...

Страница 51: ...fuse do the following Step Action 1 Press and hold the standby button until the screen turns off 2 Set the power switch a to the OFF 0 position Continued on next page WARNING Fire and electrical shock...

Страница 52: ...ge 3 Unplug the power cord b from the electric power source and the power inlet c on the warmer 4 Unscrew the cap on the fuse holder d 5 Remove the fuse e from the fuse holder 6 Replace the fuse with...

Страница 53: ...essential performance of the warmer is to not exceed an internal temperature of 160 F 71 C for blanket warmers or 150 F 66 C for fluid warmers Electromagnetic emissions The warmer is intended for use...

Страница 54: ...kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and v...

Страница 55: ...ixed RF transmitters as determined by an electromagnetic site survey1 should be less than the compliance level in each frequency range2 Interference may occur in the vicinity of equipment marked with...

Страница 56: ...ration distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 117 0 117 0 233 0 1 0 369 0 369 0 738 1 1 167 1 167 2 333 10 3 689 3 389 7 379 100...

Страница 57: ...pounds containing chloride or quaternary salts poor water quality or equipment with missing or altered serial numbers Any losses or damage resulting from malfunction including loss of contents or cons...

Страница 58: ......

Страница 59: ......

Страница 60: ...igo Products Inc 4000 S E Columbia Way Vancouver WA 98661 Phone 800 822 3501 Fax 360 696 1700 www pedigo usa com Made in the U S A Printed in the U S A Specifications are subject to change without not...

Отзывы: