Code

Refers to

Action Required

E-10

ES10

ES20

ES30

ES40

ES50

ES60

ES70

Cavity sensor

Sensor 1

Sensor 2

Sensor 3

Sensor 4

Sensor 5

Sensor 6

Sensor 7

Sensor is shorted . Software disengages heating pads . User must acknowledge error by pressing ON/OFF button . If 

error persists, a qualified service technician should test sensor. 

•  To test sensor: Detach the sensor from unit . Use an Ohm meter to measure the resistance of the sensor . Check 

sensor at 0°C (32°F) using a  container of ice water . If reading is 32 .6 KOhm ±1 .5 KOhm, replace display .  If reading 

is ±2 KOhm, replace sensor .

•  Check wires for integrity .  Check for proper and secure connections at the control and terminal block .  If necessary, 

re-secure the faulty connections .

•  Call service if error persists .

E-11

ES11

ES21

ES31

ES41

ES51

ES61

ES71

Cavity sensor

Pad sensor 1

Pad sensor 2

Pad sensor 3

Pad sensor 4

Pad sensor 5

Pad sensor 6

Pad sensor 7

Sensor is open . Software disengages heating pads . User must acknowledge error by pressing ON/OFF button . If error 

persists, a qualified service technician should test sensor. 

•  To test sensor: Detach the sensor from unit . Use an Ohm meter to measure the resistance of the sensor . Check 

sensor at 0°C (32°F) using a container of ice water . If reading is 32 .6 KOhm ±1 .5 KOhm, replace display . If reading 

is ±2 KOhm, replace sensor . 

•  Check wires for integrity . Check for proper and secure connections at the control and terminal block .  If necessary, 

re-secure the faulty connections .

•  Call service if error persists .

P130

P230

P330

P430

P530

P630

P730

Pad 1

Pad 2

Pad 3

Pad 4

Pad 5

Pad 6

Pad 7

Heating pad has not reached set-point temperature . User must acknowledge error by pressing ON/OFF 

button. If error persists, a qualified service technician should test the heater pad(s). 

•  Turn unit OFF and unplug it from AC power . 

•  Use Ohm meter to measure resistance between L:(Line) & N:(Neutral) leads of a cold heater pad . 

The Ohm readings shall be: 120V = 72 Ohm ±10% or 230V = 288 Ohm ±10% 

•  Call service if error persists .

*E-31

Cavity sensor

•  Sensor reading is above maximum allowable temperature set-point . 

Fluid Warmers, triggers at 5° over set point .  Blanket Warmers, triggers at 15° over set point

•  Call service .

P131

P231

P331

P431

P531

P631

P731

Pad sensor 1

Pad sensor 2

Pad sensor 3

Pad sensor 4

Pad sensor 5

Pad sensor 6

Pad sensor 7

Heater pad over-temp error 

•  Software disengages heating pads . 

•  User must acknowledge error by pressing ON/OFF button . 

•  Allow unit to cool . 

•  Call service if error persists .

*E-33

Cavity sensor

•  Sensor reading is above maximum allowable temperature set-point and over temp value

Fluid Warmers, triggers at 71°C (160°F) .  Blanket Warmers, triggers at 79°C (175°F)

•  Contact service

*E-50

Analog to Digital 

Convertor Error

•  Remove product and allow unit to cool down. Inspect fluid and discard if necessary. 

•  If error persists after cool down and reset, control assembly should be replaced by a qualified service technician. 

Contact service .

E-60

Real Time Clock 

Checksum Error

(Blanket Warmers Only)

Real Time Clock rechargeable battery backup has discharged . 

•  Plug unit into outlet for 30 minutes . 

•  See “Timer Control Panel” section in blanket manual to reset clock .

*E-61

Real Time Clock

(Blanket Warmers Only) 

Real Time Clock not responding . Call service if error persists .

E-62

Real Time Clock

Timer overlay is present, but no real time clock is detected . Call service .

*E-70

Pad Count Error

More heater pads detected than set for . Hold ON/OFF button for 12 seconds until display shows "PAd#” 

(# = number of pads selected [3-7]) . Press UP or DOWN arrow to adjust to correct number of pads . 

Fluid Warmers: DC250L & DC400L = 3 pads

Blanket Warmers: DC150, DC250, & EC250 = 3 pads, DC350, DC400, EC350, DC750 & EC750 = 4 pads

*E-71

Personality Error

Call service

E-80

EEPROM Error

EEPROM not responding . Call service if error persists .

*E-81

Calibration 

not locked

Call service

*E-83

EEPROM Error

Call service for help resetting the control .

E-87

EEPROM Error

Stored offsets corrupted . Offsets reset to 0 . Control may need to be recalibrated . Possible bad EEPROM . 

Call service if error persists .

E-90

Button stuck

A button has been held down for >60 seconds . Adjust control . Error will reset when the problem has 

been resolved .

E-95

Factory Test pin 

short detected .

Ensure no debris is causing a short between the test pins . If pins are good, replace control .

*E-98

Temperature 

Delta Error

(Fluid Warmers Only)  

Temperature of cavity sensors 1 and 2 differ by more than 1 .7ºC (3ºF) .

•  Remove product and allow unit to cool down . 

•  Verify that product sensor is clean and operating correctly . 

•  Press power button to clear error code .

•   If error persists, the sensor switch assembly should be replaced by a qualifi ed service technician. Contact service.

E-99

Hardware 

Over Temp

•  Inspect connections and condition of high limit bimetal thermostat .

•  If error continues call service .

*EFAn

Fan or Fan Sensor 

failure

(Fluid Warmers Only)

•  If the fan is operating, ensure that the lens on the sensor is not blocked or dirty .

•  If error persists after cleaning sensor, use an Ohm meter to ensure sensor wires are good .  A good wire will have a 

reading of <1 Ohm .

•  If wires are good, replace sensor .

•  If error persists after replacement, check sensor mount location .

•  If error still persists, replace control .

Note: 

All non-critical codes can be cleared using the ON/OFF button . Critical errors (marked with a *) can only be cleared by turning the power switch 

on the back of the unit off and allowing unit to cool .

T R O U B L E S H O O T I N G   G U I D E

Содержание P-2012

Страница 1: ...56 MM 1074 Rev 0 01 15 P 2012 P 2022 P 2032 P 2012 Shown with optional lock solid door P 2032 Shown with optional lock and left hinged door P 2022 Shown with optional solid door casters and plate p r...

Страница 2: ...eclaration 14 15 Operating Instructions Blanket Control Features 16 Operation Procedures 17 Care and Cleaning Cleaning and Preventative Maintenance Protecting Stainless Steel Surfaces 18 Cleaning Agen...

Страница 3: ...ommended environmental temperature range is 15 C to 32 C 60 F to 90 F Recommended relative humidity is above 20 non condensation E N V I R O N M E N T A L C O N D I T I O N S All Pedigo Products Inc e...

Страница 4: ...tions in this manual carefully before initiating the installation of this appliance DO NOT DISCARD THIS MANUAL This manual is considered to be part of the appliance and is to be provided to the owner...

Страница 5: ...gnal words and symbols may be used throughout this manual NOT E For equipment delivered for use in any location regulated by the following directive DO NOT DISPOSE OF ELECTRICAL OR ELECTRONIC EQUIPMEN...

Страница 6: ...uipment BS 1363 Plug UK only CEE 7 7 220 230V Plug IPX 0 Other international plugs are available Contact factory for more information Domestic ground shipping information Contact factory for export we...

Страница 7: ...V A C 60 Hz 1 ph 0 6 kW 5 0 Amps Safety Class I Equipment No Applied Parts Mode of Operation Continuous NEMA 5 15P 15A 125V Plug Hospital Grade IPX 0 220 V A C 60 Hz 1 ph 0 6 kW 2 7 Amps Type B Equipm...

Страница 8: ...1 83m est 18 0 457mm 9 1 232mm IEC Cord Inlet 42 2 1071mm 36 0 914mm 26 2 666mm with optional casters with plate 5017544 3 76mm casters 2 rigid 2 swivel w brake D I M E N S I O N S 18 5 470mm 21 9 557...

Страница 9: ...length 6 ft 1 83m est 22 8 579mm IEC Cord Inlet P 2012 continued with optional bumpers and casters 5017548 3 76mm casters 2 rigid 2 swivel w brake D I M E N S I O N S 18 5 470mm 25 5 647mm 21 2 538mm...

Страница 10: ...rd Inlet 18 0 457mm 42 3 1073mm Cord length 6 ft 1 83m est 36 0 914mm 28 8 730mm IEC Cord Inlet 26 3 669mm 9 3 235mm IEC Cord Inlet 18 0 457mm 22 0 559mm 18 5 470mm CAVITY 20 6 523mm 15 5 394mm CAVITY...

Страница 11: ...7mm 42 1 1070mm 36 3 921mm Cord length 6 ft 1 83m est 28 8 730mm IEC Cord Inlet 26 5 674mm 22 0 559mm 18 5 470mm CAVITY 20 6 523mm 26 3 669mm 15 5 394mm CAVITY 14 5 369mm INSERT 18 5 470mm 21 1 537mm...

Страница 12: ...rd Inlet 23 5 597mm 50 0 1270mm Cord length 6 ft 1 83m est 47 0 1193mm 35 6 903mm IEC Cord Inlet 28 6 726mm 12 0 305mm IEC Cord Inlet 23 5 597mm 21 0 532mm CAVITY 22 8 578mm 24 3 616mm 34 2 870mm 35 5...

Страница 13: ...4 1229mm Cord length 6 ft 1 83m est 35 6 903mm IEC Cord Inlet 24 3 616mm 28 8 732mm 21 0 532mm CAVITY 25 7 654mm 24 3 616mm 34 2 870mm 40 7 1033mm 21 0 533mm CAVITY 19 9 506mm INSERT 24 0 610mm 28 1...

Страница 14: ...2 336mm INSERT 21 4 543mm CAVITY 27 1 688mm 24 4 619mm Cord length 6 ft 1 83m est 23 5 597mm 70 0 1779mm 23 5 70 0 1779mm 27 2 691mm 35 9 911mm 19 9 506mm 128mm 21 4 543mm CAVITY 12 2 309mm IEC Cord...

Страница 15: ...779mm 28 1 715mm CAVITY 13 1 333mm INSERT 13 2 336mm INSERT 24 4 619mm 27 2 691mm 35 9 911mm 39 3 998mm 5 0 128mm 19 9 506mm 23 5 597mm 70 0 1779mm 23 5 597mm 70 0 1779mm 21 4 543mm CAVITY 12 2 309mm...

Страница 16: ...3 3 Complies Guidance and manufacturer s declaration electromagnetic immunity The unit is intended for use in the electromagnetic environment specified below The customer or the end user of this unit...

Страница 17: ...accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the unit is...

Страница 18: ...ntrol senses a temperature of 71 C 160 F or higher The alarm can be muted by pressing the ON OFF button ERROR ACKNOWLEDGEMENT To clear or acknowledge an error press the ON OFF button Press the ON OFF...

Страница 19: ...ARROW keys to change the value shown in the display The temperature set point range is 32 71 C 90 160 F 5 LOAD THE CHAMBER WITH DRY COTTON BLANKETS DO NOT WARM ITEMS CONTAINING PLASTIC RUBBER OR METAL...

Страница 20: ...grain of surface metal to avoid surface scratches Never use wire brushes metal scouring pads or scrapers to remove residue 1 Ensure that the correct Operation and Care Manual is available to all users...

Страница 21: ...ners or cleaners containing quaternary salts Never use hydrochloric acid muriatic acid on surfaces 3 Clean the interior metal surfaces of the cabinet with a damp cloth and any mild commercial detergen...

Страница 22: ...should test the heater pad s Turn unit OFF and unplug it from AC power Use Ohm meter to measure resistance between L Line N Neutral leads of a cold heater pad The Ohm readings shall be 120V 72 Ohm 10...

Страница 23: ...Error Fluid Warmers Only Temperature of cavity sensors 1 and 2 differ by more than 1 7 C 3 F Remove product and allow unit to cool down Verify that product sensor is clean and operating correctly Pre...

Страница 24: ...tory contact factory contact factory contact factory STACKING HARDWARE P 2012 or P 2022 over P 2012 or P 2022 5017272 5017272 P 2012 or P 2022 triple stacked 5017539 5017539 F U S E R E P L A C E M E...

Страница 25: ...I C E Always include both model and serial numbers in your correspondence regarding the unit Model _____________________________________ Serial Number _____________________________________ SERIAL NUM...

Страница 26: ...7998 1 P 2022 5017999 1 P 2032 5018000 1 12 PLATE LOWER HINGE RIGHT HINGED DOOR 5015523 1 LEFT HINGED DOOR 5018611 1 13 SCREW M8 x 1 25 x 20mm HEX FLG SC 27406 3 14 BOTTOM COVER P 2012 P 2022 1014270...

Страница 27: ...Model _____________________________________ Serial Number _____________________________________ SERIAL NUMBER IS REQUIRED FOR ALL INQUIRIES ELECTRICAL CHASSIS 1 2 3 5 6 4 ITEM DESCRIPTION PART QTY 1 C...

Страница 28: ...de both model and serial numbers in your correspondence regarding the unit Model _____________________________________ Serial Number _____________________________________ SERIAL NUMBER IS REQUIRED FOR...

Страница 29: ...12 P 2022 shown 2 3 1 4 5 6 7 8 12 11 10 9 ITEM DESCRIPTION PART QTY 1 OVERLAY P 2012 PE 37112 1 P 2022 PE 37111 1 P 2032 PE 37110 1 2 INTERFACE ASSEMBLY 18 5 P 2012 5018715 1 P 2022 5018717 1 INTERFA...

Страница 30: ...___ Serial Number _____________________________________ SERIAL NUMBER IS REQUIRED FOR ALL INQUIRIES SHELF P 2022 P 2032 only 36 2 3 1 4 5 6 7 8 ITEM DESCRIPTION PART QTY 1 PANEL TOP 1014609 1 2 INSULA...

Страница 31: ...MN 36956 MM 1074 Rev 0 01 15 Blanket Warming Cabinet Operation Care Manual 29 Refer to wire diagram included with the unit...

Страница 32: ...15 months from the shipping date whichever occurs first The original parts warranty on all other parts remains in effect three 3 years from installation of appliance or thirty nine 39 months from the...

Отзывы: