background image

5

Guidance and manufacturer’s declaration - electromagnetic emissions

The Model P-2010-S is intended for use in the electromagnet environment specified below.  The customer or the end user of the Model 
P-2010-S should assure that it is used in such an environment.The P-2010-S is intended for use in the electromagnetic environment specified 
below. The customer or the user of the P-2010-S  should assure that it is used in such an environment.

Immunity test

IEC 60601 test level Compliance level

Electromagnetic environment - guidance

Portable and mobile RF communications equipment should be used no closer 
to any part of the P-2010-S, including cables, than the

recommended separation distance calculated from the equation applicable to 
the frequency of the transmitter.

Recommended separation distance

Conducted RF 
IEC 61000-4-6

Radiated RF 
IEC 61000-4-3

3 Vrms 
150 kHz to 80 MHz

3 V/m 
80 MHz to 2.5 GHz

3 V

 
3 V/m

d = [3.5/3] 

P

 
d = [3.5/3] 

P  80 MHz to 800 MHz

d = [7/3] 

P  800 MHz to 2.5 GHz

where P is the maximum output power rating of the transmitter in watts 
(W) according to the transmitter manufacturer and d is the recommended 
separation distance in metres (m).  

Field strengths from fixed RF transmitters, as determined by an 
electromagnetic site survey,

a

 should be less than the compliance level in each 

frequency range.

b

Interference may occur in the vicinity of equipment marked with the following 
symbol: 

 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from 
structures, objects and people.

a

  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, 

AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment 
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the 
P-2010-S is used exceeds the applicable RF compliance level above, the P-2010-S should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as reorienting or relocating the P-2010-S.

b

 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [VI] V/m.

Guidance and manufacturer’s declaration – electromagnetic immunityRecommended separation 

distance between portable and mobile RF communications equipment and the model P-2010-S

The model P-2010-S is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.  The customer 
or the user of the Model P-2010-S can help prevent electromagnetic interference by maintaining a minimum distance between portable and 
mobile RF communications equipment (transmitters) and the Model P-2010-S as recommended below, according to the maximum output 
power of the communications equipment.

Rated maximum output  

power of transmitter

W

Separation distance according to frequency of transmitter

m

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2.5 GHz

0.01

0.117

0.117

0.233

0.1

0.369

0.369

0.738

1

1.167

1.167

2.333

10

3.689

3.689

7.379

100

11.667

11.667

23.333

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation 
applicable to the frequency of the transmitter, where P is the maximum output rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Содержание P-2010-S

Страница 1: ...OPERATION CARE MANUAL OPERATION CARE MM 1070 05 21 P R I N T E D I N U S A BLANKET WARMING CABINET P 2010 S 120V P 2010 S...

Страница 2: ...and Preventative Maintenance 9 Protecting Stainless Steel Surfaces 9 Cleaning Agents 9 Cleaning Materials 9 Care and Cleaning 10 Clean the Unit Regularly 10 Troubleshooting Guide 11 Service Parts Lis...

Страница 3: ...ent of a warranty parts and labor claim This manual must be read and understood by all people using or installing the equipment model Contact the Pedigo service department if you have any questions co...

Страница 4: ...information that is important but not hazard related 1 Pedigo blanket warmers are intended for warming cotton blankets ONLY No other use for this device is authorized or recommended 2 This device is i...

Страница 5: ...LIANCE NAMEPLATE NOTE An electrical wiring diagram for this unit can be found in a sleeve on the bottom side of the top cover of the control Safety Class I Equipment Protective Earth Ground Symbol Med...

Страница 6: ...IEC 61000 3 3 Complies Guidance and manufacturer s declaration electromagnetic immunity The Model P 2010 S is intended for use in the electromagnet environment specified below The customer or the end...

Страница 7: ...not be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength i...

Страница 8: ...tch and a temperature control knob that is adjustable from 16 C 60 F to 92 C 200 F The unit includes a digital display and a temperature display light that indicates that the cabinet has reached the d...

Страница 9: ...7 D I M E N S I O N S C L ELECTRICAL CONNECTION 1 5 38mm FROM TOP 16 25 411mm Cavity 13 25 337mm Cavity 14 365mm Cavity 21 5 546mm 20 3 516mm 20 8 528mm 27 685mm 41 1035mm...

Страница 10: ...tion and then show the temperature in C 3 Repeat to toggle to Fahrenheit Note With a power failure the control will retain the C or F setting selected by the user when power is restored 4 The temperat...

Страница 11: ...use a non abrasive scouring pad on dif cult areas and make certain to scrub with the visible grain of surface metal to avoid surface scratches Never use wire brushes metal scouring pads or scrapers to...

Страница 12: ...e on stainless steel surfaces Replace blanket support assembly 5 Clean the exterior of the cabinet with a cleaner recommended for stainless steel surfaces Spray the cleaner on a clean cloth and wipe w...

Страница 13: ...te relay Bad wire connections or open heating cable If none of the above contact Pedigo Service Department E 31 Over temperature Unit will shut down Unit has exceeded its maximum allowable set tempera...

Страница 14: ...OR HANDLE WITH STRIKER 1 HD 2007 17 HINGE SET 1 PAIR OF 2 HINGES 1 HG 2015 18 THERMOSTAT KNOB 1 KN 26568 19 TOP WELD ASSEMBLY 1 5011144 OPTIONS AND ACCESSORIES 20 CASTER STAND ASSEMBLY WITH 5 127mm CA...

Страница 15: ...13 1 2 3 4 5 6 19 16 14 12 17 11 PART NUMBERS AND DRAWINGS ARE SUBJECT TO CHANGE WITHOUT NOTICE 10...

Страница 16: ...14 REFER TO WIRE DIAGRAM UNDER TOP COVER OF P 2010 S FOR MOST CURRENT VERSION PART NUMBERS AND DRAWINGS ARE SUBJECT TO CHANGE WITHOUT NOTICE...

Страница 17: ...rs rst The parts warranty on all other parts remains in effect three 3 years from installation or thirty nine 39 months from the shipping date whichever occurs rst This warranty does not apply to 1 Ca...

Страница 18: ...MN 29621 05 21...