The essential performance of the P-2010-S is to not exceed internal temperature of 250 degrees F (+10%) at any setting.
4
G E N E R A L W A R N I N G S
The P-2010-S requires special precautions regarding EMC
and needs to be installed and put into service according to
the EMC information provided below.
Portable and mobile RF communications equipment can
affect medical electrical equipment.
The P-2010-S does not have any cables or transducers that
are detachable. The P-2010-S has a permanently attached
power cord that is ~2.5m in length.
WARNING: A risk of increased emissions or decreased
immunity may result if the power cord attached is altered or
a manufacturer supplied power cable is not used.
The P-2010-S should not be used adjacent to or stacked with
other equipment.
WARNING: observe to verify normal operation if it is
necessary to use adjacent to or stacked with other equipment.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Model P-2010-S is intended for use in the electromagnet environment specified below. The customer or the end user of the Model
P-2010-S should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Model P-2010-S uses RF energy only for it internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The Model P-2010-S is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/Flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The Model P-2010-S is intended for use in the electromagnet environment specified below. The customer or the end user of the Model
P-2010-S should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electromagnetic discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
+2 kV for power
supply lines
Mains power quality should be that of a typical
commercial or hospital environment. The
P-2010-S does not have any input/output lines.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential
mode
±2 kV common mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
<5 % UT (>95 % dip in UT)
for 0.5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 sec
<5 % UT (>95 % dip in
UT) for 0.5 cycle
40 % UT (60 % dip in
UT) for 5 cycles
70 % UT (30 % dip in
UT) for 25 cycles
<5 % UT (>95 % dip in
UT) for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of the Model P-2010-S requires continued
operation during power mains interruptions,
it is recommended that the Model P-2010-S be
powered from an uninterruptible power supply or
a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.