background image

The essential performance of the P-2010-S is to not exceed internal temperature of 250 degrees F (+10%) at any setting.

4

G E N E R A L   W A R N I N G S

The P-2010-S requires special precautions regarding EMC  
and needs to be installed and put into service according to 
the EMC information provided below.

Portable and mobile RF communications equipment can 
affect medical electrical equipment.

The P-2010-S does not have any cables or transducers that 
are detachable. The P-2010-S has a permanently attached 
power cord that is ~2.5m in length.

WARNING: A risk of increased emissions or decreased 
immunity may result if the power cord attached is altered or 
a manufacturer supplied power cable is not used.

The P-2010-S should not be used adjacent to or stacked with 
other equipment.

WARNING: observe to verify normal operation if it is 
necessary to use adjacent to or stacked with other equipment.

Guidance and manufacturer’s declaration – electromagnetic emissions

The Model P-2010-S is intended for use in the electromagnet environment specified below.  The customer or the end user of the Model 
P-2010-S should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment - guidance

RF emissions
CISPR 11

Group 1

The Model P-2010-S uses RF energy only for it internal function.  Therefore, its 
RF emissions are very low and are not likely to cause any interference in nearby 
electronic equipment.

RF emissions
CISPR 11

Class B

The Model P-2010-S is suitable for use in all establishments, including domestic 
establishments and those directly connected to the public low-voltage power 
supply network that supplies buildings used for domestic purposes.

Harmonic emissions
IEC 61000-3-2

Class A

Voltage fluctuations/Flicker emissions
IEC 61000-3-3

Complies

Guidance and manufacturer’s declaration – electromagnetic immunity

The Model P-2010-S is intended for use in the electromagnet environment specified below.  The customer or the end user of the Model 
P-2010-S should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment - guidance

Electromagnetic discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete or ceramic tile.  
If floors are covered with synthetic material, the 
relative humidity should be at least 30%.

Electrical fast transient/burst

IEC 61000-4-4

±2 kV for power supply lines

±1 kV for input/output lines

+2 kV for power 
supply lines

Mains power quality should be that of a typical 
commercial or hospital environment. The 
P-2010-S does not have any input/output lines.

Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

±1 kV differential 
mode

±2 kV common mode

Mains power quality should be that of a typical 
commercial or hospital environment.

Voltage dips, short interruptions 
and voltage variations on power 
supply input lines

IEC 61000-4-11

<5 % UT (>95 % dip in UT) 
for 0.5 cycle

40 % UT (60 % dip in UT) 
for 5 cycles

70 % UT (30 % dip in UT) 
for 25 cycles

<5 % UT (>95 % dip in UT) 
for 5 sec

<5 % UT (>95 % dip in 
UT) for 0.5 cycle

40 % UT (60 % dip in 
UT) for 5 cycles

70 % UT (30 % dip in 
UT) for 25 cycles

<5 % UT (>95 % dip in 
UT) for 5 sec

Mains power quality should be that of a typical 
commercial or hospital environment.  If the 
user of the Model P-2010-S requires continued 
operation during power mains interruptions, 
it is recommended that the Model P-2010-S be 
powered from an uninterruptible power supply or 
a battery.

Power frequency (50/60 Hz) 
magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at 
levels characteristic of a typical location in a 
typical commercial or hospital environment.

NOTE   UT is the a.c. mains voltage prior to application of the test level.

Содержание P-2010-S

Страница 1: ...OPERATION CARE MANUAL OPERATION CARE MM 1070 05 21 P R I N T E D I N U S A BLANKET WARMING CABINET P 2010 S 120V P 2010 S...

Страница 2: ...and Preventative Maintenance 9 Protecting Stainless Steel Surfaces 9 Cleaning Agents 9 Cleaning Materials 9 Care and Cleaning 10 Clean the Unit Regularly 10 Troubleshooting Guide 11 Service Parts Lis...

Страница 3: ...ent of a warranty parts and labor claim This manual must be read and understood by all people using or installing the equipment model Contact the Pedigo service department if you have any questions co...

Страница 4: ...information that is important but not hazard related 1 Pedigo blanket warmers are intended for warming cotton blankets ONLY No other use for this device is authorized or recommended 2 This device is i...

Страница 5: ...LIANCE NAMEPLATE NOTE An electrical wiring diagram for this unit can be found in a sleeve on the bottom side of the top cover of the control Safety Class I Equipment Protective Earth Ground Symbol Med...

Страница 6: ...IEC 61000 3 3 Complies Guidance and manufacturer s declaration electromagnetic immunity The Model P 2010 S is intended for use in the electromagnet environment specified below The customer or the end...

Страница 7: ...not be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength i...

Страница 8: ...tch and a temperature control knob that is adjustable from 16 C 60 F to 92 C 200 F The unit includes a digital display and a temperature display light that indicates that the cabinet has reached the d...

Страница 9: ...7 D I M E N S I O N S C L ELECTRICAL CONNECTION 1 5 38mm FROM TOP 16 25 411mm Cavity 13 25 337mm Cavity 14 365mm Cavity 21 5 546mm 20 3 516mm 20 8 528mm 27 685mm 41 1035mm...

Страница 10: ...tion and then show the temperature in C 3 Repeat to toggle to Fahrenheit Note With a power failure the control will retain the C or F setting selected by the user when power is restored 4 The temperat...

Страница 11: ...use a non abrasive scouring pad on dif cult areas and make certain to scrub with the visible grain of surface metal to avoid surface scratches Never use wire brushes metal scouring pads or scrapers to...

Страница 12: ...e on stainless steel surfaces Replace blanket support assembly 5 Clean the exterior of the cabinet with a cleaner recommended for stainless steel surfaces Spray the cleaner on a clean cloth and wipe w...

Страница 13: ...te relay Bad wire connections or open heating cable If none of the above contact Pedigo Service Department E 31 Over temperature Unit will shut down Unit has exceeded its maximum allowable set tempera...

Страница 14: ...OR HANDLE WITH STRIKER 1 HD 2007 17 HINGE SET 1 PAIR OF 2 HINGES 1 HG 2015 18 THERMOSTAT KNOB 1 KN 26568 19 TOP WELD ASSEMBLY 1 5011144 OPTIONS AND ACCESSORIES 20 CASTER STAND ASSEMBLY WITH 5 127mm CA...

Страница 15: ...13 1 2 3 4 5 6 19 16 14 12 17 11 PART NUMBERS AND DRAWINGS ARE SUBJECT TO CHANGE WITHOUT NOTICE 10...

Страница 16: ...14 REFER TO WIRE DIAGRAM UNDER TOP COVER OF P 2010 S FOR MOST CURRENT VERSION PART NUMBERS AND DRAWINGS ARE SUBJECT TO CHANGE WITHOUT NOTICE...

Страница 17: ...rs rst The parts warranty on all other parts remains in effect three 3 years from installation or thirty nine 39 months from the shipping date whichever occurs rst This warranty does not apply to 1 Ca...

Страница 18: ...MN 29621 05 21...

Отзывы: