4.3.2. Set O.C.P
(1) Enter the O.C.P setting menu:
Normal Mode
Press the System
key, enter [System → CH1 (CH2) → O.C.P].
Parallel, Series Mode
Press
the
System
key,
in
normal
status,
enter
[System
→
O.C.P];
in timer status, enter [System → Pro Set → O.C.P].
Plus-minus Mode
Press the System
key, enter [System → Positive (Negative) → O.C.P].
(2) A setting box pops up, press the
▲ / ▼ key to switch the state of the O.C.P between "ON" or "OFF", the O.C.P in
current mode and status is enabled or disabled.
Note:
In Plus-minus mode, the O.C.P status of Positive and Negative will remain consistent, you can set either of them. The
O.C.P value can be set separately.
Use the numeric keys to enter the O.C.P value in current mode and status. The maximum in Independent, Series and
Plus-minus mode is 3.15A, the maximum in Parallel mode is 6.3A. Press the
key to confirm.
4.4. Timing Output
The timing output function can preset up to 100 groups of timing parameters. When you turn on the timing output, the
instrument will output the pre-specified voltage, current in pre-specified time.
You can set special O.V.P/O.C.P for timer status; see "Over Voltage/Current Protection" for more details.
4.4.1. Enter/Exit Timer Status
Press Timer to enter/exit timer status.
The
icon identifies the system is in timer status.
4.4.2. Timer Setting
Before turning on the timing output, you should set the timer parameters, including voltage, current and output time. This
function allows up to 100 groups of timer parameters.
In timer status, press the Set key to enter/exit timer setting interface of the current mode.
Press the Mode key to switch between the corresponding timer setting interface of the four working mode.