Set all the variable controls, as follows:
Variable Controls
Setting Position
Function Select Switch (25)
Sinewave
Freq. Range Select Switch (26)
X 1 k
Freq. Control Dial (24)
1.0
Amplitude Control knob (19)
Switch off (push-in position)
Offset Control Knob (20)
Switch off (push-in position)
Symmetry Control Knob (21)
Switch off (push-in position)
Sweep Width Control Knob (22)
Switch off (push-in position)
Sweep rate Control knob (23)
Switch off (push-in position)
•
Match the output impedance between the F/G output and the unit under test by switching the impedance output
select switch (27).
•
In case that you want to measure the frequency under process, please refer to the section 5.1.4 for further
information.
5.2.3 Wave
Forms
The basic ramp signal has been generated by the OP Amp. Based on this basic signal, the squarewave has been
formed by the line receiver IC. And, the sinewave has been formed by the quardruple TR IC.
The output wave forms and phase relationship are as follows.
Note:
Under activation of the 600
Ω
impedance mode (27), the output wave forms can be normal up to 100 kHz
range.
5.2.4 Frequency
ranges
The PeakTech 4110 can supply the frequency in the 7 ranges, as shown on the following table
Frequency Range assigned to each select switch
Select Switch
Assigned Range
Maximum Output
X10
1 Hz – 10 Hz
10 Hz
X100
10 Hz – 100 Hz
100 Hz
X 1 k
100 Hz – 1 kHz
1 kHz
X 10 k
1 kHz – 10 kHz
10 kHz
X 100 k
10 kHz – 100 kHz
100 kHz
X 1 M
100 kHz – 1 MHz
1 MHz
X 10 M
1 MHz – 10 MHz
10 MHz
Note:
The output frequency can be read by multiflying figures optioned by the frequency control dial (24), and the frequency
Range select switch (26).
Ex. 1 kHz = 0.1 (at the dial 24) x 1 K (at the S/W 26)
- 28 -