Time interval A
⇒
B
Range
100 nSec to 10 Sec
LSD Display
100 nSec
Average
1, 10, 100 Intervals Averaged
Overflow Indication
The OVER LED indicator flashes, when an input exceeds, only under the LOW range set
at 10 S gate time
Operating temperature
10° C to 40° C
Storage temperature
-10°C to 50° C
Line Voltage Range
230 V AC at 50/60 Hz ± 10 % max., selectable
Power Consumption
2 W max.
Coaxial cable
(with BNC connector)
RG-58 C/U, Impedance 50
Ω
, with one (1) meter in length
5.1.2 Start-Up
•
Check conductors of the A, B and C channels in good condition
•
Connect the AC power cable to the recess power receptacle and plug it into three-contact electrical outlet
•
Turn on the Power switch of DPS
•
Turn on the Power switch of U/C located on the rear panel. In this case, all segments, PASS-ALL and UC-1300
shall be displayed as a self test.
•
Set the display Option key (28) at U/C position.
•
Set the Gate time switch (4) at 1 S position.
•
Set the Channel Select Switch (3) at Channel A position.
•
Set the Data-Hold Switch and the Attenuation Switch at the pushed-out position. Now, "GATE" "0.0000000"
"MHz" is lit up at frequency LED display.
5.1.3 Features
Channel Select Switch (3)
The Channels A and B will cover your measurement of the frequency from 5 Hz up to 100 MHz with auto range. The
Channel C will cover your measurement of the frequency from 100 MHz up to 1.3 GHz.
Gate Time Switch (4)
For a better resolution in your measurement, your selection for a proper gate time is advisable. The following table
will help you refer to the optional frequency resolution.
Frequency Resolution Table
Gate Time
100 mS
1 S
10 S
Low Range
10 Hz
1 Hz
0,1 Hz
High Range
100 Hz
10 Hz
1 Hz
Data-Hold Switch (5)
Upon activation of this mode, a read-out on the LED display will be captured and remained even after disconnecting
the test connector from a measuring point.
Attenuation Switch (2)
If input signal is greater than 700 mV, set the Attenuation Switch at the pushed-in position with connecting the input
signal to channel A only.
The triggering sensitivity of the input signal will be decrease by 1/10 (-20 dB) and reduce errors.
Reset Switch (6)
If you want to return to the normal frequency mode from the arithmetic modes, set FREQ switch at the pushed-in
position.
5.1.4
Mesuring the output frequency generated by F/G
If you want to measure the output frequency being measured by the function generator, set the display option key
(28) at F/G position.