Being in any other mode, press
【
V-SET
】
to enter the CV Mode. Press then
【
ON/OFF
】
to activate
the Load.
If the Load is deactivated, you can change the desired current with the rotating knob. Press
【
V-SET
】
once again to enter the value per key field.
Note: If the voltage on the terminals or Remote Sense is less than the set value, the CV mode
doesn't have an effect.
Note: The difference between the source voltage and the set voltage will fall onto the internal
source resistance and lead's resistance. If the difference is too high and the internal source
resistance low, high current will also flow through the Load.
6.3 Const. Power Mode (CP Mode)
In this mode, the Load will try to hold the set power value. If the voltage changes, more or less current
will be drawn.
Being in any other mode, press
【
P-SET
】
to enter the CP Mode. Press then
【
ON/OFF
】
to activate
the Load.
If the Load is deactivated, you can change the desired current with the rotating knob. Press
【
P-SET
】
once again to enter the value per key field.