Chapter 6. Maintenance 

 

This section provides some basic maintenance procedure.   

 

 

WARNING

 

The following instructions are for use by qualified service personnel only. 
To avoid electric shock, do not perform servicing other than contained in 
the operating instructions unless you are qualified to do so. 
Always unplug the equipment before performing service procedure.   

 

1. Fuse Replacement:   

 

WARNING

 

Fuse replacement should be carried out by a suitable skilled and 
competent person. 

1. Have a handy flat-bladed screwdriver. 
2. Set the equipment so its bottom is down on the work surface and it’s rear is facing you.   
3. Find the fuse holder on the rear panel.   
4. Rotate fuse holder counter-clockwise with the flat-bladed screwdriver. 
5. Pull the fuse holder away. 
6. Remove the line fuse.   
7. Reinstall: Do in reverse steps to reinstall the line fuse and fuse holder.

 

 

2. Cleaning your instrument: 

Periodically wipe the case with a damp cloth and detergent; do not use abrasives or solvents.   
 

 

Chapter 7. GPIB and RS-232 C Interface

 

 

Safety Regulations 

 
1)  

Do not touch any part of high voltage circuit or make any unnecessary high voltage measure-
ments. Do not remove the top or bottom covers of this unit unless you are a qualified service 
person. Remove the power from this unit before removing the top and bottom covers. 

2)  

When servicing this unit, the work area should be free from any electrical hazards. The floor and 
workbench where the unit is operated should be insulated and free from any exposed high 
voltage conductors. Remove any source of water or other conductive liquids in the working area. 

3)  

Connect this unit only to a 3-prong AC outlet, which conforms to electrical safety, codes. Do not 
use any adapters to connect this unit to an AC receptacle. 

4)  

Before operating or servicing this instrument, read the instruction manual and fully understand 
the operating procedures. If you are servicing the unit, check the circuit you are testing for high 
voltage. 

5)   

Do not use this device in a room alone; be sure there are other people in the vicinity of your work 
area in case of an emergency arises. Have emergency telephone numbers posted in the work 
area in case a quick response is necessary. 

 
 
 
 
 
 

Содержание 1015 GPIB

Страница 1: ...i Operation Manual Programmable Function Generator PeakTech 1015 GPIB PeakTech Spitzentechnologie die berzeugt...

Страница 2: ...ii Programmable Function Generator...

Страница 3: ...smitters vehicle radios and cellular phones generate electromagnetic radiation that could be induced into the test leads of this instrument Care should be taken to avoid such situations or alternative...

Страница 4: ...ge conductors Remove any source of water or other conductive liquids in the working area 3 Connect this unit only to a 3 prong AC outlet which conforms to electrical safety codes Do not use any adapte...

Страница 5: ...otary Dial 14 4 Quick Reference Guide 15 Chapter 5 Using this instrument 1 Equipment Needed 16 2 Preliminary setup 16 3 How to use the function wave generator 16 4 Generating a pulse waveform 18 5 Usi...

Страница 6: ...n and ranges may be controlled from a PC via the RS232 or the optional GP IB interface 8 May be used with any line voltages from 85V to 270V AC may be used with out any internal changes 9 Low Distorti...

Страница 7: ...ne Triangle Square TTL rectangular Pulse Step etc Output range Max 20 Vpp No load Max 10 Vpp into 50 Impedance 50 5 Attenuator 0 dB 20 dB DC Offset 7 5 V DC No Load 3 Wave form characteristics 1 Sine...

Страница 8: ...Use only correct size voltage and current rated fuses Test leads Inspect the test leads for damaged insulation or exposed metal Check test lead continuity Damaged leads should be replaced Chapter 2 I...

Страница 9: ...relative humidity 85 less Chapter 3 Product description In this chapter the front and rear panel buttons controls and connectors are described 1 Front Panel buttons and connectors description Program...

Страница 10: ...a digit is highlighted Pressing the Button again the cursor changes to a v symbol indicating the digit edit cursor mode Rotating the rotary dial changes the digit value 7 AMP Button Pressing this but...

Страница 11: ...NG AND TYPE DISCONNECT POWER SUPPLY BEFORE REPLACING FUSE DO NOT REMOVE COVER REFER SERVICING TO FUSE 85V 270V 50 60Hz 250V T2AL Made in Korea AC IN F U S E IEEE 488 RS 232C S N Figure 3 Rear Panel CA...

Страница 12: ...ces to the GP IB interface Note The total cable length should be less than 25m 80ft and the maximum number of device connections including controller is 15 GP IB Example Figure 6 is an example of conn...

Страница 13: ...ep settings F5 Table 1 LCD display areas Figure 5 LCD Display AMP display mode 1 Cursor 1 The cursor is displayed in the upper part of the LCD when the unit is first turned on as shown in Figure 5 It...

Страница 14: ...utton to display the selection mode cursor or b Turn the rotary dial until the cursor indicates WAVEFORM ____ c Press the button to change the cursor to the edit mode or d Rotate the rotary dial to di...

Страница 15: ...otary dial to highlight the next digit to be changed and repeat steps 1e and 1f 2 OFS DC Offset 999 to 999 The DC off offset adds a 0 to 7 5 V DC level no load to the waveform in a scale from 999 to 9...

Страница 16: ...r d Rotate the rotary dial to display the new Sweep rate 3 WIDTH Sweep width The sweep width can be adjusted from 0 05 to 19 95 Sec in a scale from 0 to 99 To change the Sweep width a Press the Button...

Страница 17: ...called from memory F2 EQUIP Displays information about your instrument including the serial number CALIBRATION date GP IB address and Firmware F3 Not used F4 LOCK Disables all front panel buttons exce...

Страница 18: ...T F4 button advances the display to the next menu screen exit returns the display to the normal screen Selects memory page address 0 to 7 FREQ WAVE FORM OUTPUT STATUS AMPL OFFSET ATTEN SYM SWEEP RATE...

Страница 19: ...s the F5 button 4 Load and display a previously saved menu page Press and release the F1 button until the desired menu page is displayed Press the F3 LOAD button 5 Press the F5 EXIT button to exit thi...

Страница 20: ...15 4 Quick Reference Guide Programmable Function Generator PeakTech 1015 GPIB...

Страница 21: ...range before turning the unit on See Figure 3 2 Connect the AC power cord to the receptacle on the rear panel 3 Turn on the power on by pressing the PWR button for 1 5 to 2 seconds 4 Let the unit War...

Страница 22: ...aveform Figure 8 Output waveform and phase relation 2 Reference supplementary explanation The following material is basic to the operation of this equipment and should be remembered when using this in...

Страница 23: ...ry of 1 1 which means the time and shape of the positive alternation is equal to the time and shape of the negative alternation This is the default state of the symmetry control in this unit The symme...

Страница 24: ...and 4 of Fig 12 3 Repetition Rate can be expressed either as a frequency or time period Repetition Rate can be measured in time on the Oscilloscope and changed to a frequency by using the formula F 1...

Страница 25: ...nction dial 4 Press the button to select the selection cursor and move the cursor to SWEEP WIDTH by rotating the rotary dial 5 Press the button to display the change cursor and rotate the rotary dial...

Страница 26: ...need the reselection to the regular frequency periodically To get the accurate frequency select the output frequency by putting the output frequency measure and the time that you have to control the...

Страница 27: ...directory named Model No automatically in Hard Disk 4 Connection of PC and Instrument Connect the RS 232 cable to the built in RS 232 connector in the Instrument and to the PC serial port 5 Communicat...

Страница 28: ...RS 232 C Interface Safety Regulations 1 Do not touch any part of high voltage circuit or make any unnecessary high voltage measure ments Do not remove the top or bottom covers of this unit unless you...

Страница 29: ...equipment setup mode 3 Press Edit button to display the movement cursor 4 Move the movement cursor to GPIB address setup field 5 Press button to change the movement mode to correction mode 6 Select t...

Страница 30: ...This includes the programming languages Quick Basic GW Basic Visual Basic Quick C Turbo C and Turbo C Communications programs such as ProComm a shareware version are also usually acceptable RS232C Sp...

Страница 31: ...ops the program 3 GPIB IEEE 488 2 Overview Introduction The instrument conforms to the Institute of Electrical and Electronics Engineers IEEE Standard 488 2 1987 The specific implementation of IEEE 48...

Страница 32: ...PIN number and signal Pin number IEEE standard Pin number IEEE Standard 1 DIO 1 13 DIO 5 2 DIO 2 14 DIO 6 3 DIO 3 15 DIO 7 4 DIO 4 16 DIO 8 5 EOI 17 REN 6 DAV 18 GND 6 7 NRFD 19 GND 7 8 NDAC 20 GND 8...

Страница 33: ...Command Description CLS Clear Status ESE GPIB ONLY Standard Event Status Enable ESE Standard Event Status Enable Query ESR Standard Event Status Register Query IDN Identification Query OPC Operation C...

Страница 34: ...age a special terminator is required CR LF CR represents carriage return and is an ASC 0D CHR 13 for basic LF represents line feed and is an ASC 0A CHR 10 for basic Command Execution examples Example1...

Страница 35: ...ymmetry 0 99 9 Symmetry SYM SYM Symmetry query SWP0 SWP0 OFF SWP1 ON SWP SWP Sweep query SWR0 05 Sweep rate 0 05 9 95 Sweep rate step 0 05 SWR SWR Sweep rate query SWW0 Sweep width 0 99 Sweep SWW SWW...

Страница 36: ...ally called a program message unit separator The followings are the examples to combine multiple commands and or queries Example1 To generate sine waveform 1kHz You would send the following program me...

Отзывы: