Page 13
Installation
Generator Environment
The generator is designed for indoor use only. It should be installed adjacent to the
application it is supplying. If this is not convenient then the unit can be sited elsewhere,
however, consideration should be made of the lengths of pipe runs as pressure drops
can result from extended runs of pipe.
Performance of the generator is affected by ambient conditions. Note should also be
taken to the proximity of Air Conditioning outlets. These can sometimes give rise to
“pockets” of air with high relative humidity. Operation of the unit within such a pocket
could adversely affect its performance. Consideration should also be given to the air
flow around the unit. We recommended that an air gap of 100mm (3.9”) should be
maintained down both sides and at the rear of the unit. Refer to the drawing on the
following page for the general dimensions of the unit.
Ensure Generator is situated in a well vetilated environment and is positioned to allow
easy disconnection from the mains supply if required
Minimum Operating Ambient Temperature:
5 °C (41 °F)
Maximum Operating Ambient Temperature:
30 °C (86 °F)