Page 10
Technical Specification
Environment
Genius XE SCI 2
Minimum Operating Ambient Temperature
5°C (41°F)
Maximum Operating Ambient Temperature
30°C (86°F)
Maximum Altitude
2000 m
Maximum Relative Humidity
80% Non-Condensing
Minimum Storage Temperature*
−20°C (−4°F)
Maximum Storage Temperature*
60°C (140°F)
*NOTE – Storage below 5°C (41°F) for long periods is not recommended. When taken out of storage the Generator
should be allowed to acclimatize at room temperature for a minimum of 3 hours before operation.
Generator Outlets
Curtain Maximum Flow^
Up to 26L/min @ 4.14bar
(0.91cfm @ 60psi)
Source Maximum Flow^
Up to 32L/min @ 7.24bar
(1.13cfm @ 105psi)
Exhaust Maximum Flow^
Up to 25L/min @ 4.83bar
(0.88cfm @ 70psi)
Particles
<0.01µm
Phthalates
NONE
Suspended Liquids
NONE
Gas Outlets
1 × 5/16" (nitrogen) 2 × ¼” BSPP (air)
Drain Outlet
1 × ¼” BSPP
^
NOTE – These performance characteristics are valid only when paired with an SCIEX approved application. Refer to the
appropriate application matrix for a list of supported products.
Electrical Requirements
Voltage
220 - 240v ±10%
Frequency
50/60 Hz
Current
9A
Input Connection
C20 Plug
Power Cord
C19 socket to local connection (10A minimum)
Pollution Degree
2
Installation / Over Voltage Category
II
General
Dimensions cm (inches) W × D × H
57 × 89.7 × 70 (22.4 × 35.3 × 27.6)
Generator Weight Kg (lbs)
103 Kg (227 lbs)
Shipping Weight Kg (lbs)
150 Kg (331 lbs)
Maximum Heat Output
5545 BTU/hr