Air Purity

The Solaris XE generator should be connected to an air supply that, as a minimum, 

meets ISO8753-1:2010 Class 1.4.1

Class 1 Particulate

In each cubic metre of compressed air, the particulate count should not exceed 20,000 

particles in the 0.1 - 0.5 micron size range, 400 particles in the 0.5 - 1 micron range and 

10 particles in the 1 - 5 micron size range.

Class 4 Water

A pressure dew point (PDP) of +3°C or better is required and no liquid water is allowed.

Class 1 Oil

In each cubic metre of compressed air, not more than 0.01mg of oil is allowed. This is a 

total level for liquid oil, oil aerosol and oil vapour.

Connecting to the application

Select the ¼”- 6mm push fitting from the fittings kit and connect to  the generator AIR 

IN and NITROGEN ports on the rear panel. Using suitable 6mm tubing, connect the 

outlet of the generator to the inlet on the application.

Please refer to the Tubing Lengths section, to ensure correct tubing is being used.

Once the tubing is connected to the application, please ensure that it 

is thoroughly checked for being leak-tight. Even the slightest leak in 

the gas supply between the generator and the application can lead to 

a reduction in efficiency.

Tubing Lengths

The diameter of the tubing which will be connected to the gas outlet is 

important and is determined by the length of tubing required. Failure 

to follow these recommendations could lead to pressure between 

generator and application.

 

< 10 meters:  

Use 

¼

”/

3

16

” (

¼

” O/D, 

3

16

” I/D) P.T.F.E. tubing. 

 

> 10 - 40 meters:  Use 

3

8

”/ 

5

16

” (

3

8

” O/D, 

5

16

” I/D). Tubing and fittings not supplied  

    in 

the 

fittings 

kit.

 

> 40 metres:  

Please contact Peak Scientific with the relevant distance and  

 

 

 

 

we will calculate the flow resistance and the tubing size required.

A combination of 

¼

”/

3

16

” and 

3

8

”/

5

16

” tubing may be used to ensure that there is no large 

diameter tubing within the lab (i.e. for the first 20 meters from the generator use 

3

8

”/

5

16

” 

and the final 10 meters to the application use 

¼

”/

3

16

” tubing). Keep the connections and 

bends to a minimum.

CAUTION

CAUTION

Содержание Solaris XE

Страница 1: ...Solaris XE User Manual Copyright Peak Scientific Instruments Ltd 2020 Solaris XE User Manual Rev 5 RSID 2302 EN 21 07 2020...

Страница 2: ...Kit Contents 10 Unpacking Instructions 10 Installation 11 Generator Environment 11 Generator Overview 12 Solaris XE General Dimensions 12 Solaris XE Rear Connections 13 Electrical Connection 14 Air P...

Страница 3: ...ntended for end users and has been written as a reference document where you can skip to the relevant information Users can refer to the contents page to find the relevant information Please review ea...

Страница 4: ...ality of the foregoing that any defect is not the direct or indirect result of lack of repair and or servicing incorrect repair and or servicing use of wrong materials and or incorrect spare parts 2 A...

Страница 5: ...ted conditions are fully understood or met A CAUTION notice denotes a hazard It calls attention to an operating procedure process or similar which if not correctly performed or adhered to could cause...

Страница 6: ...10 Electrical Equipment for measurement control and laboratory use Electromagnetic Compatibility Directive 2014 30 EU EN 61326 1 2013 Electrical Equipment for measurement control and laboratory use FC...

Страница 7: ...onfirm that as a manufacturer of articles we conform to the European Parliament and Council Directives 2011 65 EU for the original 6 restricted substances listed in the Restriction of Hazardous Substa...

Страница 8: ...Flow For Max Outlet Flow up to 100 L min Minimum Air Inlet Quality ISO 8573 1 2010 Class 1 4 1 Generator Outlets Maximum Output Pressure 116 psi 7 99 bar variable Maximum Outlet Flow Rate up to 35 L m...

Страница 9: ...k Partner from where the unit was purchased Follow the unpacking instructions posted on the side of the crate It will require two people to remove the unit from the shipping crate and to manoeuvre the...

Страница 10: ...Key x 1 10 No 8 Wall Plug x 4 11 No 8 Screw x 4 All of the generators output ports are located on the rear panel of the unit Unpacking Instructions The unit weight constitutes a two person lift and a...

Страница 11: ...ths of pipe runs as pressure drops can result from extended runs of pipe The performance of the generator like all sophisticated equipment is affected by ambient conditions Note should also be taken t...

Страница 12: ...Page 12 Generator Overview Solaris XE General Dimensions Horizontal postioning on wall 343 mm 13 50 655 mm 25 78 156 mm 6 14...

Страница 13: ...ts are connected to correct application Connections should only be carried out by trained personnel The Generator should be switched off prior to maintenance and cleaning Power Switch Power Inlet Nitr...

Страница 14: ...se check that your electrical supply falls within the specified range see below EARTH GROUND E Green Yellow or Green LIVE L Brown or Black Neutral N Blue or White Electrical requirements are 100 230VA...

Страница 15: ...tion to ensure correct tubing is being used Once the tubing is connected to the application please ensure that it is thoroughly checked for being leak tight Even the slightest leak in the gas supply b...

Страница 16: ...ive an indication of internal pressure in the system The LED s on the front panel will illuminate YELLOW Turn the air supply on The fascia will illuminate GREEN when adequate system pressure has reach...

Страница 17: ...the regulator at the front of the generator and viewed on the adjacent gauge To access the regulator unclip the fascia from the unit Adjust the gauge to the required pressure To ensure a smooth press...

Страница 18: ...uirements Service Schedule Purchase Interval Component Visit Qty 12 months Solaris Nitrogen XE Annual Maintenance Kit www peakscientific com ordering 1 Please note Only Peak Scientific approved compon...

Страница 19: ...to install and setup your generator IQ OQ Certified assurance for applications requiring documented qualification Complete plan Swift response by a Peak Service Engineer within 72 hours planned preve...

Страница 20: ...e cloth prior to use Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the generator Under no circumstances should any solvents or abrasive cle...

Страница 21: ...rd plug Contact your service provider The fascia LED s have not illuminated but the power switch is illuminated Contact your service provider The fascia Yellow illumination does not change to green Ch...

Страница 22: ...ic com For generators that remain unregistered after 1 month from the shipment date the warranty will be considered active from the date of factory dispatch Complete Plan only Go Online or Complete an...

Страница 23: ...Page 23...

Страница 24: ...ponse service to our customers With Peak Protected your laboratory s productivity becomes our top priority To discuss Peak Protected generator cover and payment options speak to your local Peak Repres...

Отзывы: