Inlet Filter / Separator Elements
These should be changed at intervals as indicated below. In addition filter bowls should be cleaned and
the operation of the auto-drains should be checked.
The Generator MUST be de-pressurised prior to attempting to remove ANY filter bowl. Failure to
do this may cause injury.
Inlet Filter
This should be changed at 6-month intervals. Part Number is 02-4187 and the filter is located as shown.
Remove the cover by rotating it anti-clockwise ¼ turn. The element can then be removed. Re-fitting is the
reverse procedure.
Inlet Filter
(02-4187)