PEAK
Scientific Instruments Ltd
ZA300 Zero Air Gas Generator
Instructions for use Manual
30
th
March 2005
ZA300 issue 3
18
7
Technical Specifications
General Details
Minimum Operating Ambient Temperature
5
o
C (41
o
F)
Maximum Operating Ambient Temperature
45
o
C (113
o
F)
Inlet Conditions
(Free of oil and bulk moisture)
Minimum Air Inlet Pressure
90psig
Maximum Air Inlet Pressure
130psig
Minimum Air Inlet Flow Rate
Gas outlet flow & Purge flow
36LPM
Outlet Gas
Maximum Pressure Drop (Outlet-Inlet)
δ
P
N/A
Maximum Gas Outlet Pressure
100 psig (6.8 Barg)
Maximum Gas Outlet Flow for specified hydrocarbons
30 Litres/min (ATP)
Pressure Dewpoint
-30
o
C (-22
o
F)
Start up time for dewpoint
8 hrs
Particles
0.01um
Hydrocarbon concentration (as methane)
<0.1ppm
Start up time for hydrocarbon concentration
45 minutes
O
2
level
20%
Electrical Requirements
@230V ac (50 Hz)
8.9 AMPS
Electrical Connection
IEC-Euroconnector
General
Dimensions W x D x H cm
43 x 41x 88
Weight Kg
64
Shipping Dimensions W x D x H cm
100 x 57 x 67
Shipping Weight Kg
91