Page 7

6.

5.

Insert the 

6mm Blanking Plug (02-1268)

 from the fi ttings kit into the 

Drain

 port at 

the rear of the MS Table.

OPTIONAL:

 To attach the 

water bottle mount

, locate the two mounting holes, 

towards the back of the MS Table on the left side. Using the screws and washers 

provided attach the water bottle mount 

in the order of: Water bottle mount, Plain 

Washer (00-1044), Spring Washer (00-1044) then M5 Screw (00-5476)

. Do this for 

both mounting holes.

02-1268

00-1040

00-5476

00-1044

Содержание MS Table Series

Страница 1: ...MS Table Installation Guide Rev 1 PN IG 0012 21 09 16 MS Table All Models Installation Guide...

Страница 2: ...or more extensive user operation directions we recommend you download the User Manual from the website Please visit www peakscientific com downloads to download the full User Manual for your gas gener...

Страница 3: ...CAUTION notice denotes a hazard It calls attention to an operating procedure process or similar which if not correctly performed or adhered to could cause damage to the Generator or the Application D...

Страница 4: ...ush Fit Fitting 02 4600 x1 UK Mains Cable C13 04 1028 x1 US Mains Cable C13 6 15 04 1029 x1 EU Mains Cable C13 04 1038 x1 US Mains Cable C13 5 15 04 1039 x6 M5 Plain Washer 00 1040 x6 M5 Spring Washer...

Страница 5: ...Page 5 Unpack the MS Table from the shipping crate and position on a flat surface in desired area 1 2 To lock the levelling feet in place firstly slacken off the 17mm top nut 0 0...

Страница 6: ...foot a further 1 full turn before tightening the locking nut to lock the foot in place For non gas versions of the MS Table proceed to step 8 or for instructions on fitting the optional water bottle m...

Страница 7: ...the water bottle mount locate the two mounting holes towards the back of the MS Table on the left side Using the screws and washers provided attach the water bottle mount in the order of Water bottle...

Страница 8: ...compressed air inlet bulkhead located at the rear of the unit 04 1028 04 1029 04 1038 04 1039 7 Select the appropriate mains cable from the fittings kit and plug the IEC 60320 C19 cable into the main...

Страница 9: ...60Hz OR 230VAC 50 60Hz Plug the mains cable into the corresponding 110VAC 50 60Hz OR 230VAC 50 60Hz single phase power supply CAUTION ensure the correct cable has been selected for your electrical sup...

Страница 10: ...Page 10 12 11 After 30 minutes of running the generator can be disconnected from the air supply The 6mm blanking plug can now be removed from the Drain port...

Страница 11: ...our instrument Re connect the compressed air supply to the Air inlet Connect 6mm PE Tubing 00 1282 to the Drain outlet Fit the other end of the tubing to a suitable drain connection or container The c...

Страница 12: ...Page 12 CONGRATULATIONS Your Peak Scientific MS Table is now fully installed operational and ready to supply gas on demand to your instrument 15...

Страница 13: ...stance away from the instrument the generator can be placed still using 6mm 4mm internal diameter or 3 16 internal diameter tubing is 10m For further information relating to tubing lengths please refe...

Страница 14: ...ic com For generators that remain unregistered after 1 month from the shipment date the warranty will be considered active from the date of factory dispatch Complete Plan only Go Online or Complete an...

Страница 15: ...or MS Table Installation Guide Rev 1 PN IG 0012 21 09 16 Peak Scientific has highly trained fully certified Field Service Engineers located in over 20 countries across every continent around the world...

Отзывы: