Page 14
16.
15.
The outlet pressure reading on the display
will gradually increase
to the factory set
pressure of 100 psi (6.9 bar).
GENIUS XE 35
S/N: XXXXXXXXXXXX
Nitrogen:
XXXX
psi
The generator will now initialise. To ensure that the correct voltage is being supplied
to the generator, the generator will show the supply voltage on its front display.
The
voltage reading should be checked prior to the initial purge of the system
. The
supply voltage will be displayed on the touch screen when the generator is switched
on.
If the reading is 219V or less, then we would highly recommend fitting a Dual Tap
Transformer 06-3200 (XE 35) or 06-3210 (XE 70).
This can be ordered directly from
Peak Scientific.
For Genius XE 35 (120V) voltage should not be less than 114V.
[Initialisation...]
Model : GENIUS XE 35
Software : v0.00
Voltage : 230V RMS
Initialisation Time : 00:00
Continue...