Page 19
Tubing Lengths
The diameter of the tubing which will be connected to the gas outlet is
important and is determined by the length of tubing required. Failure
to follow these recommendations could lead to pressure between
generator and application.
< 10 meters:
Use 6/4 (6mm O/D, 4mm I/D) P.T.F.E. tubing.
> 10 - 40 meters: Use 10/8 (10mm O/D, 8mm I/D). Tubing and fittings not
supplied in the fittings kit.
> 40 metres: Please contact Peak Scientific with the relevant distance andwe
will calculate the flow resistance and the tubing size required.
A combination of 6/4 and 10/8 tubing may be used to ensure that there is no large
diameter tubing within the lab (i.e. for the first 20 meters from the Generator use 10/8
and the final 10 meters to the application use 6/4 tubing). Keep the connections and
bends to a minimum.
The imperial equivalents are: 6/4 = 1/4” O/D, 3/16” I/D.
10/8 = 3/8” O/D, 5/16” I/D.