WARNING
Electrical Connection
Connect the generator to an appropriate 110 or 230 volt single-phase supply, refer to
the generator serial plate for input specification and ensure your supply matches the
requirements.
If the appropriate power cord is not supplied; a new plug, rated to at least 12 amps, can
be fitted by a qualified electrician.
This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE
EARTHED. Before connecting the unit to the mains supply, please
check the information on the serial plate. The mains supply must be of
the stated AC voltage and frequency.
EARTH/GROUND (E):-
Green & Yellow or
Green
LIVE (L):-
Brown
or
Black
Neutral (N):-
Blue
or
White
Electrical requirements for 230V generators is 195V - 253V 50/60Hz. For 195V - 219V a
06-3200 Dual Tap Transformer is required.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment maybe impaired.
Drain Connection
Please ensure that the drain port at the rear of the compressor is led to a suitable
connection or container, It should be noted that the generator will expel considerable
amounts of water from this port. If a container is used it should be emptied at regular
intervals.
Note: The container must not have an airtight seal.
WARNING