background image

GeniusN118LA-N418LA User Manual Rev 2 PN UM-NX18LA 18/11/21

Genius N118LA & N418LA

User Manual

Содержание Genius N118LA

Страница 1: ...GeniusN118LA N418LA User Manual Rev 2 PN UM NX18LA 18 11 21 Genius N118LA N418LA User Manual...

Страница 2: ...11 Genius N118LA N418LA 11 Unpacking 12 Fittings Kit Contents 13 Installation 14 Generator Environment 14 Generator Overview 15 General Dimensions 15 Rear Connections 16 Electrical Connection 17 Drain...

Страница 3: ...document where you can skip to the relevant information Users can refer to the contents page to find the relevant information Please review each of the following sections carefully Thank you for sele...

Страница 4: ...designed as a plug and play system Economical More cost effective than any other gas supply method Convenient Gas on demand no hazards no need to worry about running out of gas The Genius N118LA N418L...

Страница 5: ...d air or inlet air supply to the generator must comply with ISO 8573 1 2010 Class 1 2 1 at all times d your air compressor dryer filtration and oil removal systems must be deemed suitable for use by P...

Страница 6: ...be deleted and the rest shall remain unaffected continuing to be in full force and effect Delivery of Warranty Service 12 Subject to clause 13 and a Peak being notified by you within the duration of t...

Страница 7: ...ted conditions are fully understood or met A CAUTION notice denotes a hazard It calls attention to an operating procedure process or similar which if not correctly performed or adhered to could cause...

Страница 8: ...ative requirements Low Voltage Directive 2014 35 EU EN 61010 1 2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use Electromagnetic Compatibility Directive 2014...

Страница 9: ...al Equipment Safety Regulations 2016 SI 2016 1101 as amended BS61010 1 2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use The Electromagnetic Compatibility Re...

Страница 10: ...Guidance Notes PDF produced by the Department for Business Innovation and skills for the UK and here for Europe All PEAK products that are subject to the WEEE directive are compliant with the WEEE mar...

Страница 11: ...ets Maximum Gas Output Pressure 6 9 bar 100 psi Maximum Outlet Flow Rate 18 L min 0 64 cfm Particles 0 01 m Phthalates NONE Suspended Liquids NONE Gas Outlets 1 x BSPP Drain Outlets 1 x BSPP Pressure...

Страница 12: ...e Peak Partner from where the unit was purchased Follow the unpacking instructions posted on the side of the crate It will require two people to remove the unit from the shipping crate and to manoeuvr...

Страница 13: ...the Fittings Kit are as follows 1 Teflon Tubing x 3m 2 6mm PE Tubing x 3m 3 6mm Teflon Tubing x 3m 4 Compression Fitting x 1 5 6mm Push Fit Fitting x 2 6 UK Mains Power Cable x 1 7 EU Mains Power Cab...

Страница 14: ...s affected by ambient conditions Note should also be taken to the proximity of Air Conditioning outlets These can sometimes give rise to pockets of air with high relative humidity Operation of the uni...

Страница 15: ...15 Generator Overview General Dimensions The generator must always be placed on a flat level surface Failure to do so will affect the performance of the generator WARNING 400 mm 15 7 700 mm 27 5 710 m...

Страница 16: ...Page 16 Rear Connections Compression Fitting 6mm Push Fit Fitting Mains Cable...

Страница 17: ...ly must be of the stated AC voltage and frequency EARTH GROUND E Green Yellow or Green LIVE L Brown or Black Neutral N Blue or White Electrical requirements for 230V generators is 195V 253V 50 60Hz Fo...

Страница 18: ...will continue throughout the operation of the Generator When the system has been operated for a period of 30 minutes all the internal pipe work and storage tanks will have been purged with Nitrogen T...

Страница 19: ...Tubing and fittings not supplied in the fittings kit 40 metres Please contact Peak Scientific with the relevant distance andwe will calculate the flow resistance and the tubing size required A combina...

Страница 20: ...essors The rate at which they cycle will be dependent on the gas required to satisfy the demand of the application If the application demands the maximum gas flow of the Generator the compressor duty...

Страница 21: ...xtreme conditions that can affect the duty cycling of the compressor These extremes include very high ambient temperatures low supply voltages a very high flow requirement and locations of high altitu...

Страница 22: ...to install and setup your generator IQ OQ Certified assurance for applications requiring documented qualification Complete plan Swift response by a Peak Service Engineer within 72 hours planned preve...

Страница 23: ...e cloth prior to use Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the generator Under no circumstances should any solvents or abrasive cle...

Страница 24: ...ned on switch in the up position Reconnect power cord Contact your service provider Compressors are running but pressure is not building Check pressure gauge is showing normal pressure Contact your se...

Страница 25: ...ic com For generators that remain unregistered after 1 month from the shipment date the warranty will be considered active from the date of factory dispatch Complete Plan only Go Online or Complete an...

Страница 26: ...ry leading rapid response service to our customers With Peak Protected your laboratory s productivity becomes our top priority To discuss Peak Protected generator cover and payment options speak to yo...

Отзывы: