Page 25
Troubleshooting
Problem
Possible Solution
The Generator will not switch on and the
power switch does not illuminate.
• Ensure power cord is plugged into the
Generator and that the power socket is
turned on.
• Check the fuse in the power cord plug.
• Contact your service provider.
The Generator will not switch on but the
power switch is illuminated.
• Disconnect power cord from the rear of
the Generator. Open the left hand panel
are check that both circuit breakers are
turned on (switch in the up position).
Reconnect power cord.
• Contact your service provider.
Compressors are running but pressure is not
building.
• Check pressure gauges are showing
normal pressure.
• Contact your service provider.
The mass spec is reporting low pressure.
• Contact your service provider.
Yellow “SERVICE” LED on front panel is on
constantly.
• A compressor(s) due for service. Contact
your service provider.
• Refer to Service Indication section of this
manual for further information.
*Yellow “SERVICE” LED on front panel is
flashing.
• A compressor(s) is overdue for service.
Contact your service provider urgently.
• Refer to Service Indication section of this
manual for further information.
Yellow “HIGH DUTY” LED on front panel is on
constantly.
• A compressor(s) has not cycled for a
period of 8 hours. Refer to the High Duty
Indication section of this manual.
• Ensure ambient temperatures are within
the specification.
• Ensure there are no leaks between the
Generator and the mass spec.