Page 14
Installation
Generator Environment
The generator is designed for indoor use only. It should be installed adjacent to
the application(s) it is supplying. If this is not convenient then the unit can be sited
elsewhere, however, consideration should be made of the lengths of pipe runs as
pressure drops can result from extended runs of pipe.
Performance of the generator (like all sophisticated equipment) is affected by ambient
conditions. Note should also be taken to the proximity of Air Conditioning outlets.
These can sometimes give rise to “pockets” of air with high relative humidity. Operation
of the unit within such a pocket could adversely affect its performance. Consideration
should also be given to the air flow around the unit. It is recommended that an air gap
of 75mm (3”) should be maintained down both sides and at the rear of the unit. Please
refer to the drawing below for the general dimensions of the unit.
Minimum Operating Ambient Temperature:
5 °C (41 °F)
Maximum Operating Ambient Temperature:
35 °C (95 °F)