8 CAMWARE SOFTWARE
Explanation:
•
Off
: MCP gating is disabled,
Intensifier Voltage
is continously
on; there is no contribution of the MCP to the overall extinction
ratio. Maximum fps can only be achieved with MCP gating
Off
•
On:
MCP gating is enabled; MCP
Intensifier Voltage
is
switched off after the end of the photocathode exposure and
reactivated immediately after the sCMOS sensor readout is
done. With this setting, the MCP additionally contributes to the
overall extinction ratio. Reactivation of the
Intensifier Voltage
takes an extra 4 ms; this mode slows down the maximum
achievable framerate.
Example timing & frame rate for MCP Gating Mode “on”:
Type
Value
Type of phosphor
P46
Exposure time
10 µs
Configured phosphor decay time
10 µs
Readout time for a single image @full resolution
9.5 ms
MCP reactivation time
4 ms
Frame time
13.5 ms
Resulting frame rate @ full resolution
1
13.5
𝑚𝑚𝑚𝑚 ≈
74
𝑓𝑓𝑓𝑓𝑚𝑚
Phophor Decay
Additional configurable sCMOS exposure time to detect the
“afterglow” of the phosphor screen. Each phosphor has a specific
phosphor decay time, depending of the used material (P43, P46)
See detailed explanation in chapter
Phosphor
Phosphor decay typ. to
…10% …1%
Typical efficiency
P43
1 ms
4 ms
100%
P46
0.2 … 0.4 µs
2 µs
30%