MOST 20-2B
7
PACIFIC CONSOLIDATED INDUSTRIES
350003 Rev D
SECTION IV
4-1 FILLING
INSTRUCTIONS
WARNING:
Fill the MOST with high purity
oxygen conforming to OXYGEN, USP. Ensure
no flames or sparking devices are within a 50 ft
radius of MOST or the High Pressure Supply (No
Smoking).
Filling of the MOST should be conducted
outdoors. Ensure High Pressure Oxygen Supply
Line and Oxygen Input Port connections are
clean and debris free.
Use only authorized cleaning procedures.
Connections contaminated with oils or
hydrocarbons can easily initiate a fire.
Always open the oxygen valves slowly to
minimize the compression heating problem.
WARNING:
Composite cylinders require
Hydrostatic Testing at five (5) year intervals.
Check Hydrostatic Test plate prior to filling to
ensure cylinders testing is current.
1.
Ensure hands are clean or wear lint free clean gloves.
2.
Ensure High Pressure Oxygen Source is SHUT OFF.
3.
Ensure Venting Valve is in VENT position.
See Figure 4-2.
4.
Disconnect all hoses from the Oxygen Outlet
Connectors.
5.
SLOWLY OPEN the Cylinder Shutoff Valves.
See Figure 4-1.
6.
Remove protective plug on Oxygen Input Port.
7.
Attach high-pressure oxygen source hose between the
source device and the MOST oxygen input port (use
Adaptor fitting if required).
8.
SLOWLY turn the Venting Valve on MOST
to the CLOSED position (See Figure 4-2).
9.
VERY SLOWLY OPEN the High Pressure Oxygen
Supply valve on source device. The MOST will begin
filling at a rate of approximately 200 psi/minute.
10.
Do not exceed fill pressure of 2,250 psi as indicated
by the cylinder contents gauge.
11.
When filling is complete, CLOSE the High Pressure
Oxygen Supply Valve at source device.
12.
Switch Venting Valve (V-3) to VENT position.
(Oxygen will briefly vent through Vent Port)
13.
Remove High Pressure Oxygen Supply Hose from
Oxygen Input port (C-4).
14.
Switch Venting Valve to CLOSE position.
15.
Replace protective cap on Oxygen Input port.
16.
SLOWLY CLOSE the Cylinder Shutoff Valves.
(Shutoff Valves)
Cylinder Shutoff Valves V-1 & V-2
Figure 4-1
Note: