The sensor unit must be installed in the room to be inspected, such as a room where CO

2

 gas bottles are 

stored. The LC display shows the CO

concentration in the ambient air as well as the temperature. When 

the green LED glows constantly, the power supply of the device is granted. 
 

 If 

“ESC” is indicated in the LC display of the remote display unit, you must adopt protective 

measures before entering the room where the sensor unit has been installed. 

Check the room’s 

ventilation and find out whether there is a CO

leak, wearing protective clothing / equipment. As soon as 

the CO

concentration in the ambient air has reached 

the second alarm threshold, “ESC” appears in the 

display of the sensor unit as well as in the display of the remote display unit

. The yellow LED “FLT” 

(standing for fault) flashes. Use the 

„ReFactSet“ function to reset the gas leak detector to factory defaults. 

The “ESC” indication should now disappear. Alternatively, you can remove the mains adaptor and plug it 
back into the socket. 
 
The sensor unit has the following functions: 
 

DIAG

   

Communication test between sensor unit and remote display unit 

CAL

 

 

Calibration of gas detector 

ReFactSet

 

Reset to factory default settings 

 
You can set two alarm thresholds: 

AL1 

AL2 

1. Alarm threshold 

 
Selectable threshold values: 0.5 %; 1 %; 1.5 %; 2 

(Factory default setting: 1.5 %) 
 
When the CO

2

 concentration exceeds the first 

threshold value, the red AL1 LED (LED 1) will flash, 
the buzzer will sound and the potential-free contact 
will close if the red and blue cables are connected 
or closes when the white and blue cables are 
connected. 
 
The AL1 LED will stop flashing and the buzzer will 
stop sounding as soon as the CO

2

 concentration 

falls below the first  alarm threshold value. 

2. Alarm threshold 

 
Selectable threshold values: 1.5 %; 2 %; 2.5 %; 3 
%; 3.5 %; 4 % 
(Factory default setting: 3 %) 
 
When the CO

2

 concentration exceeds the second 

threshold value, the red AL1 and AL2 LEDs will 
flash at the same time, the time between the 
buzzer intervals becomes shorter and „ESC“ 
appears in the LC display. 
 
When the CO

2

 concentration falls below the second 

alarm threshold, even if it falls below the first alarm 
threshold, the red AL1 and AL2 LEDs will keep 
flashing and the buzzer will sound until you restart 
the device. To restart the gas meter, press the 
“Reset“ key (F) or disconnect the mains plug from 
the socket and plug it back in. 

 
 
 

 

Содержание PCE-WMM 50

Страница 1: ...iter FL 33458 USA From outside US 1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Busine...

Страница 2: ...n 6 3 1 Device 6 3 2 Display 9 4 Getting started 10 4 1 Preparation 10 4 2 Power connection 11 5 Operation 12 5 1 Further measuring functions 12 5 2 Settings 13 6 Maintenance 14 6 1 Repair cleaning an...

Страница 3: ...tightly with the elastic plug lock so that it can only be removed using mechanical tools If you wish to connect a fan to the potential free contacts of the sensor unit make sure that the fan has its o...

Страница 4: ...250 V AC max 2 A Alarm thresholds Alarm 1 0 5 1 1 5 2 Alarm 2 1 5 2 2 5 3 3 5 4 Menu items Function test of connection between sensor unit and remote display unit calibration switching between C and F...

Страница 5: ...nal digital display to measure temperature in C and F LEDs for operation alarms 1 2 and monitoring of voltage supply Display elements Function test Reset menu selection Enter shifting between C and F...

Страница 6: ...tential free contact which can for instance activate a fan which reduces the increased CO2 concentration in the room concerned Sensor unit A LC display B Yellow LED error indication C Red LED 2 AL2 D...

Страница 7: ...between sensor unit and remote display unit CAL Calibration of gas detector ReFactSet Reset to factory default settings You can set two alarm thresholds AL1 AL2 1 Alarm threshold Selectable threshold...

Страница 8: ...te display unit O Green LED power supply P Red LED 1 AL1 Q Red LED 2 AL2 R Yellow LED error indication S LC display T Enter key U Mode key V RJ45 connection Output W RJ45 connection for sensor unit In...

Страница 9: ...d LED 1 flashes the buzzer sounds and the potential free contact closes if the red and blue cables are connected or closes if the white and blue cables are connected Second alarm threshold If the CO2...

Страница 10: ...2 concentration before entering the room where the sensor unit is located The connection between both devices can be checked by using the DIAG function which is described in the further course of this...

Страница 11: ...blue cables are connected the contact closes B blue cable R red cable W white cable 1 power connection fan 2 fan 4 2 Power connection Connect the mains adaptor to the wall socket Attach the elastic sa...

Страница 12: ...larms are triggered early 5 1 Further measuring functions DIAG function The communication between the sensor unit and the remote display unit should be checked on a regular basis by using the DIAG fun...

Страница 13: ...actory default settings are restored after the confirmation 5 2 Settings Lock unlock The gas detector has an automatic lock function to prevent the settings from being changed unintentionally If you w...

Страница 14: ...e key The following options are available 1 5 2 2 5 3 3 5 4 Select the desired option and press the Enter key to confirm 6 Maintenance 6 1 Repair cleaning and maintenance Repairing Do not try to repai...

Страница 15: ...ubber cap on the gas input M to prevent water from getting into the device Maintenance The communication between the sensor unit and the remote display unit should be checked on a regular basis by usi...

Страница 16: ...f this does not help please contact our customer support Er7 Internal data transfer error Er7 flashing LED FLT flashing buzzer makes noise Er7 flashing LED FLT flashing buzzer makes noise Unplug the p...

Страница 17: ...use them or give them to a recycling company which disposes of the devices in line with law If you have any questions please contact PCE Instruments 9 Contact If you have any questions about our rang...

Отзывы: