c
- ON/OFF
-
ESC
d
- Change Setting
- Calibration
e
- Menu
-
Confirm
f
- Store Data
g
- Increase Symbolized Number
- Backlight Switch
h
- Re-read Data
1. Switch on
Press
key to turn on the gauge. The gauge will consecutively display
- Logo
- Serial No. and software version then the gauge will enter into measuring mode and display
- 0.0mm/0.00mm(0.00IN/0.000IN depending on the last used)
- Velocity rate and selected probe
- Location Lumber
2. Backlight on / off
Press
key, the backlight is on. Press
key again, the backlight is off.
3. Measurement
There are three ways to go into measuring mode:
- Switch on the gauge;
- Press power key
escape from MENU;
- Measurement.
4. Calibration
When initially switch on the gauge or get an error value instead of 4.00mm (when velocity rate is 5920 m/s) if
take measurement of test block attached on the gauge (which correct value should be 4.00mm) or other
standard test block, the gauge needs to be calibrated simply by pressing
key. The gauge will display same
information and operate according to this information, then the gauge will automatically calibrate.
5. Set-up
Press
key, the gauge will display:
1. Measurement
2. Velocity
Rate
3. Probe
Setting
4. Resolution
5.
Memory
6.
Limitation
7. Average
Measurement