Conductivity Calibration 

  

PCE-PWT 10 pocket salinity tester allows up to 3 points calibration in the conductivity mode. To 

ensure higher accuracy, we recommend that you perform 3 points calibration or select a standard 

value close to the sample value you are measuring. The tester will automatically detect these 

conductivity standard solutions and prompt the user to calibrate the meter. When the calibration 

is done, all new calibration values will automatically override existing data. The following table 

shows acceptable conductivity range of calibration solution for each measuring range.  

 

 

MEASURING RANGE   CALIBRATION 

SOLUTION 

RANGE  

DEFAULT  

0~20µS/cm  

7~17µS/cm  

10µS/cm  

20~200µS/cm  

70~170µS/cm  

84µS/cm  

200~2000µS/cm  

700~1700µS/cm  

1413µS/cm  

2~20mS/cm  

7~17mS/cm  

12.88mS/cm  

 

Ensure that you use fresh conductivity standard solution during the calibration. Do not reuse 

calibration solutions as it may be contaminated and affect the calibration and accuracy of 

measurement. 

  

 

 

Содержание PCE-PWT 10

Страница 1: ...ter Analysis Meter User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last cha...

Страница 2: ...int Calibration 8 10 2 Point Calibration 9 11 3 Point Calibration 10 12 Temperature Calibration 10 13 Switching the Measurement Mode 11 14 Measurement 11 15 Hold Function 11 16 Electrode Care and Main...

Страница 3: ...ons The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should on...

Страница 4: ...0 pocket salinity tester has a succinct membrane keypad names and symbols which describe each of the function key controls KEY FUNCTION MEAS HOLD Power the tester ON OFF Freezes the measured values on...

Страница 5: ...er is in the measurement mode CAL Indicates tester is in the calibration mode SETUP Indicates tester is in SETUP mode HOLD Indicates the measured value has been frozen ATC Indicates temperature compen...

Страница 6: ...ff to conserve energy To disable the auto off function please read the SETUP MENU section PARAMETER DESCRIPTION OPTION S DESCRIPTION DEFAULT Calibration Point 1 point 2 points 3 points Temperature Uni...

Страница 7: ...tester goes to temperature unit selection mode the display shows o C UNIT 3 Press CAL key to select the temperature unit o C or o F and press ENTER key to confirm The tester goes to temperature calibr...

Страница 8: ...function is enabled When the auto off function is enabled if you do not press any keys for 8 minutes the tester will automatically turn off to conserve energy 7 Press CAL key to enable or disable the...

Страница 9: ...te the meter When the calibration is done all new calibration values will automatically override existing data The following table shows acceptable conductivity range of calibration solution for each...

Страница 10: ...immediately shows current calibration standard e g 1413 S cm 1 4 Press ENTER key the default calibration value begins flashing 1 5 Press CAL key to set each digit press ENTER key to confirm When the...

Страница 11: ...de into the calibration solution the tester automatically shows current calibration standard e g 12 88mS cm 4 Press ENTER key the default calibration value begins flashing 5 Press CAL key to set each...

Страница 12: ...12 Temperature Calibration During the measurement if the temperature reading displayed differs from that of an accurate thermometer you need to calibrate the tester 1 Press and hold the CAL key for 3...

Страница 13: ...two data hold modes When the Auto Hold function is enabled the tester will automatically sense a stable endpoint reading and freeze it HOLD indicator appears on the display If the Auto Hold function...

Страница 14: ...tly push the module into the tester Twist the electrode collar clockwise until it is tight Installation is complete 18 Replacing the Batteries If the battery indicator is disappearing during use you w...

Страница 15: ...0 00 10 00ppt Accuracy 1 F S Resolution 0 01ppt Conductivity Range 0 20 00 200 0 2000 S cm 20 00mS cm Accuracy 1 F S Resolution 0 01 0 1 1 Calibration Points 1 to 3 points Calibration Solutions 10 S...

Страница 16: ...neral Business Terms which you can find here https www pce instruments com english terms 24 Disposal For the disposal of batteries in the EU the 2006 66 EC directive of the European Parliament applies...

Страница 17: ...583 975 114 Fax 1 561 320 9176 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce americas com info industrial needs com info pce italia it www pce instruments com us www pce instruments com english w...

Отзывы: