5.2 

Automatic power-off 

As per factory default, the device turns off automatically after 15 minutes of inactivity. 
This feature can be adjusted by the user as follows: 

 

With the device turned off, press the mA%/V and the power button simultaneously to 
turn the device on. 

 

Release the power button as soon as the display shows all of the content. 

 

The  device  is  now  in  calibration  mode  and  the  symbols  AP-XX  are  shown  on  the 
display. 

 

Use the 

▲ or ▼ buttons to toggle between AP-OFF and AP-ON. 

 

Press the 100%/START button to store the changes. 

 

Press the power button to exit. 

5.3 

Device output 

The device produces a DC current from its appropriate output terminal 

OUTPUT

set by the 

user or it can simulate transmitter output. 
 

 

Caution

 

Do not apply voltage to the output terminal during the operation. If any improper voltage is applied 
to the output terminal, it will cause damage to the internal circuitry. 
 
 

Function 
Operation 

% Operation 

Display 

Setting Range 

DCI  20mA 

20 mA 

 % 

00.000 mA 
-025.00 mA % 

00.000

22.000  mA 

-025.00

112.50 mA % 

 

5.3.1 

Current output 

 

Insert one end of the test lead to the + mA 

– output jack

OUTPUT

of the device and 

connect the other end with the input of the device as shown in Figure 5.1. 

 

Press the output/input toggle button (3) so that the  OUTPUT symbol appears on the 
display. The device is now in an output state. 

 

Press the mA%/V unit select button (4) so the device displays mA or mA%; where 0% 
is 4mA and 100% is 20mA. 

 

Use the 

◄ or ► buttons to select the output digits. 

 

Use the 

▲ or ▼ buttons to change the value of the set digits. The value can be rounded 

or decremented automatically. Press and hold the button for one second to change the 
value continuously. 

 

Содержание PCE-LOC 10

Страница 1: ...oop Calibrator User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change...

Страница 2: ...3 3 2 LCD layout 4 4 Device Maintenance 4 4 1 General maintenance 4 4 2 Battery replacement 5 4 3 Fuse replacement 6 5 Operating Instructions 6 5 1 Power on off 6 5 2 Automatic power off 7 5 3 Device...

Страница 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Страница 4: ...4V DC 25 mA 10 Voltage measurement Area Resolution Accuracy 0 2 28V 1 mV 0 02 of meas 2 mV Current measurement Area Resolution Accuracy 1 22 mA 0 001 mA 0 02 of meas 4 A Current measurement Loop curre...

Страница 5: ...evice layout 1 LCD Display 6 25 single step Auto ramp 2 Power 7 0 100 of setting Auto ramp start 3 Output Input toggle 8 Output value set 4 mA unit select 9 Input digit select 5 Single step Auto toggl...

Страница 6: ...input and output values ON OFF These symbols denote the status of the input output signals These symbols denote the high and low speed of the auto ramp and auto step ramp 4 Device Maintenance This se...

Страница 7: ...olarity of the batteries with the polarity on the device Remove the batteries if the device is not to be used for a prolonged period of time Dispose of the old batteries according to local regulations...

Страница 8: ...with a screwdriver Replace the blown fuse s Reinstall the cover Reinstall the device protector Figure 3 2 Fuse replacement 5 Operating Instructions 5 1 Power on off Press the power button to turn the...

Страница 9: ...ution Do not apply voltage to the output terminal during the operation If any improper voltage is applied to the output terminal it will cause damage to the internal circuitry Function Operation Opera...

Страница 10: ...vice displays mA or mA Use the or buttons to change the output value in 25 increments where 0 is 4mA and 100 is 20mA Press the 25 RAMP button again to exit the step current output 5 3 3 Current Output...

Страница 11: ...ol is displayed press the 100 START button to start setting the waveform output Press the 100 START button again the output will be paused at the current value and the OFF symbol will be displayed Pre...

Страница 12: ...V 28 V 20 mA 00 000 mA 25 00 mA 0 000 V 1 000 22 000 mA 31 25 112 50 mA 0 2000 28 000 V 5 4 1 DC current measurement Insert one end of the test lead to the mA jack INPUT of the device and connect the...

Страница 13: ...symbol appears on the display The device is now in an input state Press the mA V unit select button 4 so the device displays V The device now begins to measure ON and the measured result are now show...

Страница 14: ...ge Output Range Resolution Accuracy Remark DCA 20mA 0 000 22 000mA 0 001mA 0 05 set value 4uA Max load 1K at 20mA Simu transmitter absorption current 20mA 0 000 22 000mA 0 001mA 0 05 set value 4uA Max...

Страница 15: ...isposal of batteries in the EU the 2006 66 EC directive of the European Parliament applies Due to the contained pollutants batteries must not be disposed of as household waste They must be given to co...

Страница 16: ...583 975 114 Fax 49 0 2903 976 99 9971 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce instruments com info pce instruments co uk info pce italia it www pce instruments com deutsch www pce instrumen...

Отзывы: